Purpose: To show the feasibility of 3D-printed fixation masks for whole brain radiation therapy in a clinical setting and perform a first comparison to an established thermoplastic mask system.

Methods: Six patients were irradiated with whole brain radiotherapy using individually 3D-printed masks. Daily image guidance and position correction were performed prior to each irradiation fraction. The vectors of the daily position correction were compared to two collectives of patients, who were irradiated using the standard thermoplastic mask system (one cohort with head masks; one cohort with head and neck masks).

Results: The mean systematic errors in the experimental cohort ranged between 0.59 and 2.10 mm which is in a comparable range to the control groups (0.18 mm-0.68 mm and 0.34 mm-2.96 mm, respectively). The 3D-printed masks seem to be an alternative to the established thermoplastic mask systems. Nevertheless, further investigation will need to be performed.

Conclusion: The prevailing study showed a reliable and reproducible interfractional positioning accuracy using individually 3D-printed masks for whole brain irradiation in a clinical routine. Further investigations, especially concerning smaller target volumes or other areas of the body, need to be performed before using the system on a larger basis.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC8166668PMC
http://dx.doi.org/10.1007/s11548-021-02393-2DOI Listing

Publication Analysis

Top Keywords

thermoplastic mask
12
3d-printed masks
12
3d-printed fixation
8
fixation masks
8
masks brain
8
established thermoplastic
8
patients irradiated
8
individually 3d-printed
8
position correction
8
cohort head
8

Similar Publications

Objective: The primary aim of this investigation is to assess the effectiveness of implementing an innovative immobilization approach, spec ically the utilization of personalized open-face masks in combination with styrofoam fixation, for head and neck cancers receiving radiotherapy. The study seeks to evaluate the influence of this method on improving patients' precision in positioning and their overall comfort during the treatment process, in addition to exploring its potential capacity to mitigate the occurrence of anxiety and depression in this patient population.

Methods: A prospective, randomized controlled trial was undertaken to investigate the comparative efficacy of two immobilization approaches for the radiotherapy treatment of head and neck cancers.

View Article and Find Full Text PDF

Background: Implementing the thermoplastic mask as a treatment option in Gamma Knife radiosurgery (GKRS) has madethe discussion of the relative benefits of frame- or mask-based fixation for treatment a frequent occurence. Based on a case review, the authors discuss factors that need to be understood and considered by the treating team.

Observations: The authors describe the case of an 81-year-old patient with pleomorphic dermal sarcoma that extended to and infiltrated the dura mater, who underwent extensive resection requiring a skin flap.

View Article and Find Full Text PDF

This study compared three Brainlab thermoplastic masks-Cranial 4pi basic, stereotactic (Close Mask V2), and open-face-to see how well they limited head movement during Stereotactic Radiosurgery (SRS). Using a head phantom, we tested rotational movements (pitch, yaw, and roll) and measured displacements with the ExacTrac system. The open-face mask had the smallest mean displacements (pitch: 0.

View Article and Find Full Text PDF

Objective: To evaluate the safety and efficacy of a modern linear accelerator with non-invasive thermoplastic mask-based immobilization as an alternative treatment for medically refractory tremor.

Background: Essential and Parkinsonian tremors are prevalent movement disorders impacting millions globally. Although Gamma Knife radiosurgery remains a conventional treatment for these tremors, it demands invasive frame immobilization, long-treatment time, & periodic Co-60 source replacements.

View Article and Find Full Text PDF

3D-printed boluses in radiation therapy are of increasing interest for enhancing treatment precision and patient comfort. A comprehensive clinical validation of these boluses remains to be established. This study aims to confirm the effectiveness of a 3D-printed bolus through a proof-of-concept comparative validation, by implementing in a clinical setting a bolus made of PLA and designed to ensure uniform dose coverage for a case in the eye region.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!