Background: Venous congestion can be quantified by central venous pressure (CVP) and its monitoring is crucial to understand and follow the hemodynamic status of patients with cardio-respiratory diseases. The standard technique for CVP measurement is invasive, requiring the insertion of a catheter into a jugular vein, with potential complications. On the other hand, the current non-invasive methods, mainly based on ultrasounds, remain operator-dependent and are unsuitable for use in the home environment. In this paper, we will introduce a novel, non-invasive device for the hospital, office and home assessment of CVP.

Methods: After describing the measurement concept, we will report a preliminary experimental study enrolling 5 voluntary healthy subjects to evaluate the VenCoM measurements' repeatability, and the system's capability in measuring small elicited venous pressure variations (2 mmHg), as well as an induced venous hypertension within a pathological range (12÷20 mmHg).

Results: The experimental measurements showed a repeatability of ±1mmHg. The VenCoM device was able to reliably detect the elicited venous pressure variations and the simulated congestive status.

Discussion And Conclusion: The proposed non-invasive VenCoM device is able to provide a fast and repeatable CVP estimate, having a wide spectrum of potential clinical applications, including the monitoring of venous congestion in heart failure patients and in subjects with renal and hepatic dysfunction, as well as pulmonary hypertension (PH) that can be extended to pneumonia COVID-19 patients even after recovery. The device needs to be tested further on a large sample size of both healthy and pathological subjects, to systematically validate its reliability and impact in clinical setting.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC8128499PMC
http://dx.doi.org/10.2147/MDER.S307775DOI Listing

Publication Analysis

Top Keywords

venous pressure
16
novel non-invasive
8
non-invasive device
8
central venous
8
hospital office
8
venous congestion
8
elicited venous
8
pressure variations
8
vencom device
8
venous
7

Similar Publications

Background: Pulmonary arterial hypertension (PAH) is characterized by elevated pulmonary artery pressure and vascular resistance, leading to systemic venous hypertension and potential right heart failure. These elevated pressures can extend to ocular veins, resulting in complications such as central retinal vein occlusion (CRVO). This case report highlights a rare instance of CRVO combined with cilioretinal artery occlusion (CilRAO), an uncommon ocular manifestation associated with PAH.

View Article and Find Full Text PDF

Carotid body tumors (CBTs), rare neuroendocrine neoplasms near the carotid bifurcation, are mostly asymptomatic but may cause discomfort and autonomic dysfunction. Computed tomography angiography (CTA) is used for diagnosis, eliminating the need for a biopsy to avoid the risk of hemorrhage. Surgical excision is the preferred treatment, while radiotherapy is an option when surgery is impractical.

View Article and Find Full Text PDF

Objectives: Femtosecond laser image guided high precision trabeculotomy (FLigHT) is a novel open-angle glaucoma treatment. The procedure non-invasively creates aqueous humor (AH) drainage channel from the anterior chamber (AC) to Schlemm's canal (SC) through the trabecular meshwork (TM) to decrease intraocular pressure (IOP). The purpose of this study was to develop a 3D finite element model (FEM) of the FLigHT procedure and to simulate clinical results for different drainage channel cross-sectional areas.

View Article and Find Full Text PDF

Objective: This study aims to investigate the clinical application value of the central venous-arterial carbon dioxide partial pressure difference (Pv-aCO2) in postoperative cognitive dysfunction(POCD) in patients with acute aortic dissection.

Methods: A retrospective analysis was conducted on the general data of 236 patients. Blood gas samples were collected from the arterial and venous lines at various time points during the surgery, including before and after the initiation of cardiopulmonary bypass (CPB), immediately after CPB initiation, before and after deep hypothermic circulatory arrest, 30 min after rewarming, and 5 min before weaning from CPB.

View Article and Find Full Text PDF

Background Cardiorenal syndrome (CRS) refers to the bidirectional interactions between the acutely or chronically dysfunctioning heart and kidney that lead to poor outcomes. Due to the evolving literature on renal impairment and heart failure with preserved ejection fraction (HfpEF), this review aimed to highlight the pathophysiological pathways, diagnosis using imaging and biomarkers and management of CRS in patients with HFpEF. Further studies are needed to validate the use of novel biomarkers, especially for early diagnosis and prognostication.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!