We performed 84 intraocular pressure measurements with the Oculab Tono-Pen, Goldmann applanation tonometer, and pneumatic tonometer in 47 eyes that had undergone pars plana vitrectomy and gas-fluid exchange. Measurements made by using the Tono-Pen were accurate when compared to those made by Goldmann tonometry (mean difference, 0.74 mm Hg). In a subset of eyes with increased intraocular pressure (greater than or equal to 25 mm Hg), the Tono-Pen provided measurements similar to those made by Goldmann applanation tonometry (P greater than .60), with only three of 39 readings (8%) underestimating the Goldmann pressure by more than 3 mm Hg. Pneumatic tonometry significantly underestimated the intraocular pressure in eyes with increased pressure (P less than .001), with 20 of 39 readings (51%) underestimating the Goldmann pressure by 5 mm Hg or more.