AI Article Synopsis

  • Commercial large-bore suction catheters and makeshift catheters using endotracheal tubes (ETTs) were tested against standard Yankauer catheters for suction efficiency.
  • The study aimed to compare the suction rates using these devices with different fluids, measuring performance across various viscosities.
  • Results showed that the commercial large-bore catheter consistently provided faster suction rates than both the Yankauer and the ETT combination, indicating its superiority for clinical use.

Article Abstract

Background: Commercial large-bore suction catheters and makeshift large-bore suction catheters with an endotracheal tube (ETT) attached to a meconium aspirator have been shown to have superior suction rates compared with a standard Yankauer.

Objectives: To compare suction rates between a commercial large-bore suction catheter and a makeshift endotracheal tube-meconium aspirator (ETTMA) combination.

Methods: We measured suction rates between a standard bulb-tip Yankauer catheter (Model: K86; Cardinal Health, Waukegan, IL), a commercial large-bore catheter (Part: 43200; The Big Stick®, SSCOR, Inc., Sun Valley, CA), and an 8.0-mm standard ETT connected to a meconium aspirator (NeotechTM Products, Inc., Valencia, CA; Ref: N0101 Clear) with high-, medium-, and low-viscosity fluids. Median suction rates were calculated with interquartile ranges. The relative differences with 95% confidence intervals (CI) between the Yankauer and both the large-bore catheter and the ETTMA combination were calculated using a linear mixed-effects model.

Results: Each device was trialed five times with each of the three different viscosity fluids for a total of 45 trials. Overall, suction rates were faster for the large-bore suction catheter compared with the Yankauer (relative difference 22 mL/s; 95% CI 17-28) and ETTMA (8 mL/s; 95% CI 5-10). The large-bore catheter had consistently faster suction rates compared with the Yankauer and ETTMA combination across all fluid viscosities.

Conclusions: The commercial large-bore suction catheter had faster suction rates than the makeshift ETTMA combination when compared with the standard Yankauer.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.jemermed.2021.02.028DOI Listing

Publication Analysis

Top Keywords

suction rates
32
large-bore suction
28
commercial large-bore
20
suction
14
suction catheter
12
large-bore catheter
12
ettma combination
12
large-bore
10
rates
8
rates standard
8

Similar Publications

Objective: Percutaneous vegetation debulking has been reported to treat tricuspid valve infective endocarditis (TVIE), but data on feasibility compared with conventional surgical strategies are limited. We aimed to compare short-term outcomes of suction debulking with partial venovenous bypass to conventional open surgery in this population.

Methods: This was a single-center, retrospective study that included all patients with isolated TVIE who underwent suction debulking with partial venovenous bypass or tricuspid valve surgery between January 2010 and December 2022.

View Article and Find Full Text PDF

The highest unqualified cleaning rate of suction-type lumen instruments is a major challenge for a central sterile supply department (CSSD). However, A few comprehensive studies have analyzed the main factors affecting cleaning quality. In response, this study aimed to explore the current state and the factors affecting the cleaning quality of reused suction-type metal lumen instruments in CSSD.

View Article and Find Full Text PDF

Objectives: To study the long-term outcomes of the flexible and navigable suction ureteral access sheath (FANS) in flexible ureteroscopy (FURS). FANS has well-established efficiency and 30-day perioperative safety; however, its influence on pelvicalyceal and ureteric anatomy remains to be investigated.

Methods: This was a prospective study of patients with normal renal anatomy who underwent FURS with FANS for renal stones from April 2023 to August 2024 in 16 centers worldwide.

View Article and Find Full Text PDF

Background And Objective: Lipedema is a commonly underdiagnosed chronic condition. This study aimed to evaluate liposuction techniques for lipedema by conducting a systematic review and presenting our experience. A case series study and a comprehensive review were conducted.

View Article and Find Full Text PDF

Introduction: Tranexamic acid (TXA) is an antifibrinolytic drug commonly used in total knee arthroplasty (TKA). Intravenous (IV) and topical TXA therapy have been extensively studied and shown to reduce blood loss, length of hospital stay, and blood transfusion rates following TKA. Despite the extensive literature regarding IV and topical TXA in orthopedics, there is a current dearth of studies analyzing oral usage.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!