Comparison of Transmittance and Reflectance Pulse Oximetry in Anesthetized Dogs.

Front Vet Sci

Clinic of Small Animal Medicine, Centre for Clinical Veterinary Medicine, Ludwig-Maximilians-University (LMU) Munich, Munich, Germany.

Published: April 2021

The tongue is the standard site for placement of a pulse oximeter probe but is difficult to access during certain procedures such as dental and ophthalmic procedures and computerized tomography of the head. The aim of this study was to evaluate the performance of a new-generation reflectance pulse oximeter using the tail and tibia as sites for probe attachment. A total of 100 client-owned dogs that underwent anesthesia for various reasons were premedicated with butorphanol ( = 50; 0.2 mg/kg; group BUT) or butorphanol and dexmedetomidine ( = 50; 5 μg/kg; group DEX), administered intravenously. Anesthesia was induced with propofol and maintained with sevoflurane. A transmittance pulse oximeter probe was placed on the tongue and served as the reference standard. A reflectance probe was randomly placed on the tail base or the proximal tibia, and the position changed after testing. Signals from three consecutive measurements were obtained at each position. Failure was defined as "no signal," "low signal," or a pulse difference >10/min compared with the ECG heart rate. Data were analyzed using chi-square test, Wilcoxon matched-pair signed-rank test, and Bland-Altman analysis. < 0.05 was considered significant. In both groups (BUT and DEX), failure rate was higher when the tibia and tail were used as probe sites compared with the tongue. In both groups, the failure rate was higher for the tibia than for the tail. Dexmedetomidine-induced vasoconstriction increased failure rate at all probe positions. The tail base, but not the tibia, is an acceptable position for reflectance pulse oximeter probes in dogs. The tongue remains the probe site of choice, if accessible.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC8119740PMC
http://dx.doi.org/10.3389/fvets.2021.643966DOI Listing

Publication Analysis

Top Keywords

pulse oximeter
16
reflectance pulse
12
failure rate
12
dogs tongue
8
oximeter probe
8
tail base
8
rate higher
8
higher tibia
8
tibia tail
8
probe
7

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!