Comparative efficacy and safety of the Captus device for inferior vena cava filter retrieval.

Clin Imaging

Department of Radiology, Division of Interventional Radiology, University of Colorado Anschutz Medical Campus, Mail Stop # L954, 12401 E. 17th Avenue, Aurora, CO 80045, United States of America. Electronic address:

Published: September 2021

Purpose: Retrievable inferior vena cava filters (IVCF) have been increasingly used for mechanical pulmonary embolism prophylaxis since their development. The Captus Vascular Retrieval System (Avantec Vascular, Sunnyvale, California) is a new device developed for retrieval of IVCF. This study compared the safety and efficacy of the new Captus device against the existing EnSnare Endovascular Snare System (Merit Medical, South Jordan, Utah) for IVCF retrieval.

Methods: Patients undergoing IVCF retrieval at a single institution between July 2015 and July 2020 were retrospectively identified. All adult patients (>18 years) undergoing filter retrieval with either Captus or Ensnare were included. Technical success and complications were compared by device. A complexity score was assigned to each case to adjust for selection bias. Logistic regression was used to model the association between device type and primary technical success.

Results: 99 IVCF retrievals met inclusion criteria, 59 with Captus and 40 with Ensnare. The majority of the cohort consisted of low complexity cases (n = 51, 86% Captus versus n = 31, 78% Ensnare; p = 0.28). Technical success for low and medium complexity retrievals was 88% and 62% with Captus and 96% and 33% with Ensnare. There was no significant association between device type and technical success, adjusting for case complexity (Captus OR 0.55, 95% CI 0.08-2.72, p = 0.49). There were no device-related complications.

Conclusion: No statistically significant difference in device technical success or complications between the Ensnare and Captus devices for uncomplicated IVCF retrieval.

Precis: The Captus Vascular Retrieval System is a new device for IVC filter retrieval which has similar technical success to the existing EnSnare.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.clinimag.2021.04.024DOI Listing

Publication Analysis

Top Keywords

technical success
20
filter retrieval
12
captus
10
device
8
captus device
8
inferior vena
8
vena cava
8
captus vascular
8
vascular retrieval
8
retrieval system
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!