Model-Based Optimization of Spinal Cord Stimulation for Inspiratory Muscle Activation.

Neuromodulation

Department of Biomedical Engineering, University of Michigan, Ann Arbor, MI, USA; Biointerfaces Institute, University of Michigan, Ann Arbor, MI, USA; Department of Anesthesiology, University of Michigan, Ann Arbor, MI, USA. Electronic address:

Published: December 2022

Objective: High-frequency spinal cord stimulation (HF-SCS) is a potential method to provide natural and effective inspiratory muscle pacing in patients with ventilator-dependent spinal cord injuries. Experimental data have demonstrated that HF-SCS elicits physiological activation of the diaphragm and inspiratory intercostal muscles via spinal cord pathways. However, the activation thresholds, extent of activation, and optimal electrode configurations (i.e., lead separation, contact spacing, and contact length) to activate these neural elements remain unknown. Therefore, the goal of this study was to use a computational modeling approach to investigate the direct effects of HF-SCS on the spinal cord and to optimize electrode design and stimulation parameters.

Materials And Methods: We developed a computer model of HF-SCS that consisted of two main components: 1) finite element models of the electric field generated during HF-SCS, and 2) multicompartment cable models of axons and motoneurons within the spinal cord. We systematically evaluated the neural recruitment during HF-SCS for several unique electrode designs and stimulation configurations to optimize activation of these neural elements. We then evaluated our predictions by testing two of these lead designs with in vivo canine experiments.

Results: Our model results suggested that within physiological stimulation amplitudes, HF-SCS activates both axons in the ventrolateral funiculi (VLF) and inspiratory intercostal motoneurons. We used our model to predict a lead design to maximize HF-SCS activation of these neural targets. We evaluated this lead design via in vivo experiments, and our computational model predictions demonstrated excellent agreement with our experimental testing.

Conclusions: Our computational modeling and experimental results support the potential advantages of a lead design with longer contacts and larger edge-to-edge contact spacing to maximize inspiratory muscle activation during HF-SCS at the T2 spinal level. While these results need to be further validated in future studies, we believe that the results of this study will help improve the efficacy of HF-SCS technologies for inspiratory muscle pacing.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC8876828PMC
http://dx.doi.org/10.1111/ner.13415DOI Listing

Publication Analysis

Top Keywords

spinal cord
24
inspiratory muscle
16
lead design
12
hf-scs
10
cord stimulation
8
muscle activation
8
muscle pacing
8
inspiratory intercostal
8
contact spacing
8
neural elements
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!