Background: Epidural catheters are routinely placed for many surgical procedures and to treat various pain conditions. Known complications arising from epidural catheter equipment malfunction include epidural pump failure, epidural catheter shearing, epidural catheter connector failure, epidural filter connector cracking, and loss-of-resistance syringe malfunction. Practitioners need to be aware of these potentially dangerous complications and take measures to mitigate the chances of causing significant patient harm. We report on the complete breakage of an epidural filter connector during epidural bolus administration of local anesthetic by hand with a syringe.

Case Presentation: A B. Braun Perifix® epidural catheter was placed in a 73-year-old male scheduled for radical prostatectomy. During the operation, a continuous infusion of local anesthetic was administered through the epidural catheter in addition to general endotracheal anesthesia. At the conclusion of surgery and after extubation, the patient endorsed incisional pain. The epidural filter connector broke in half as a bolus of local anesthetic was administered by hand with a syringe. The local anesthetic sprayed widely throughout the room as the fragmented epidural filter connector became a projectile object that recoiled and struck the patient.

Conclusions: This incident placed the patient and surrounding healthcare providers at substantial risk for injury and infection from the fractured epidural filter connector becoming a projectile object and from the local anesthetic spray. The most plausible cause of this event was from a large amount of pressure being applied to the filter connector. This may have occurred by excessive force being applied by hand to the syringe, by the presence of a clogged filter, or by the catheter being kinked or blocked proximal to the filter. Being aware of this deleterious complication and potentially modifying existing epidural bolus techniques, such as using smaller syringes with less applied force and checking all epidural components vigilantly prior to and during bolus administration, can help anesthesia providers deliver the safest possible care to patients with epidural catheters.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC8114482PMC
http://dx.doi.org/10.1186/s12871-021-01372-zDOI Listing

Publication Analysis

Top Keywords

filter connector
28
epidural filter
24
local anesthetic
24
epidural catheter
20
epidural
16
bolus administration
12
filter
9
connector
8
administration local
8
epidural catheters
8

Similar Publications

Introduction: A specialized device equipped with a sharp blade filter has been developed to enable more efficient purification of a micronized cellular adipose matrix (MCAM) containing stem cells. The aim of this study is to compare the characteristics and functions of the population of stromal cells (mSVF) and cultured cells (mASCs) purified using this device with those of cSVF and cASCs obtained through conventional enzymatic purification.

Methods: Cell viability, proliferation capacity and yield were assessed.

View Article and Find Full Text PDF

The impacts of effluent discharge on receiving waterbodies have been a research hotspot. Nonetheless, limited information is available on the microbial community assembly patterns in the hyporheic zones (HZ) responding to the changes in the microenvironments, e.g.

View Article and Find Full Text PDF

In dual-band RF front-end systems, to transmit different frequency signals in different paths, each path requires the power to be divided along two transmission channels. In such systems, a circuit is created in which the input ports of power dividers with different frequency bands are connected to the output ports of a diplexing circuit in a cascade form. These circuits often contain different band filters in their schemes and have a complicated design.

View Article and Find Full Text PDF

Brain computer interfaces (BCIs) provide clinical benefits including partial restoration of lost motor control, vision, speech, and hearing. A fundamental limitation of existing BCIs is their inability to span several areas (> cm) of the cortex with fine (<100 μm) resolution. One challenge of scaling neural interfaces is output wiring and connector sizes as each channel must be independently routed out of the brain.

View Article and Find Full Text PDF

A capillary-based immunofluorescence sensor was developed and incorporated in a flow injection analysis system. The light-guiding capillary was illuminated axially by a 473 nm/5 mW solid state laser through a tailored optofluidic connector. High sensitivity of the system was achieved by efficiently collecting and detecting the non-guided fluorescence signal scattered out along the wall of the capillary.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!