Background: Magnetic resonance imaging (MRI) is an examination method for noninvasive soft tissue imaging without the use of ionizing radiation. Metallic implants, however, may pose a risk for the patient and often result in imaging artifacts. Due to the increasing number of implants, reducing these artifacts has become an important goal. In this review, we describe the risks associated with implants and provide the background on how metal-induced artifacts are formed. We review the literature on methods on how to reduce artifacts and summarize our findings.
Method: The literature was searched using PubMed and the keywords "MRI metal artifact reduction", "metallic implants" and "MRI artefacts/artifacts".
Results And Conclusion: The MRI compatibility of implants has to be evaluated individually. To reduce artifacts, two general approaches were found: a) parameter optimization in standard sequences (echo time, slice thickness, bandwidth) and b) specialized sequences, such as VAT, OMAR, WARP, SEMAC and MAVRIC. These methods reduced artifacts and improved image quality, albeit at the cost of a (sometimes significantly) prolonged scan time. New developments in accelerated imaging will likely shorten the scan time of these methods significantly, such that routine use may become feasible.
Key Points: · Metallic implants may pose a risk for patients and often cause artifacts.. · Imaging artifacts can be reduced by parameter optimization or special sequences.. · Metal artifacts are reduced with a lower TE, smaller voxel size, larger matrix, and higher bandwidth.. · SPI, STIR, VAT, SEMAC, MAVRIC, and MAVRIC-SL are specialized MR sequences that can reduce artifacts further..
Citation Format: · Peschke E, Ulloa P, Jansen O et al. Metallic Implants in MRI - Hazards and Imaging Artifacts. Fortschr Röntgenstr 2021; 193: 1285 - 1293.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1055/a-1460-8566 | DOI Listing |
J Orthop Sci
January 2025
Department of Pediatric Orthopedics, Hunan Provincial People's Hospital, The First Affiliated Hospital of Hunan Normal University, Changsha, Hunan, 410013, China.
Backgroud: Medial humeral epicondyle fracture is a prevalent type of upper limb fractures in pediatric patients. This study aims to compare the follow-up clinical results and complications in 30 children with medial epicondyle fractures who were treated with either metal screws or absorbable screws at our hospital.
Methods: A retrospective review was conducted on 30 children with medial humeral epicondyle fractures, who were divided into two groups: Metal group (18 children) underwent fixation using metal screws, while Absorbable group (12 children) received absorbable screws between January 2016 and June 2024.
Sci Rep
January 2025
Department of Oral and Maxillofacial Surgery, University Medical Center Groningen, University of Groningen, Hanzeplein 1, P.O. Box 30.001, 9700 RB, Groningen, The Netherlands.
In cases of large mandibular continuity defects resulting from malignancy resection, the current standard of care involves using patient-specific/custom titanium reconstruction plates along with autogenous grafts (fibula, scapula, or iliac crest segments). However, when grafts are not feasible or desired, only the reconstruction plate is used to bridge the gap. Unfortunately, metal osteosynthesis and reconstruction plates, including titanium, exhibit adverse effects such as stress-shielding and limitations in accurate postoperative irradiation (especially with proton-beam therapy).
View Article and Find Full Text PDFACS Nano
January 2025
School of Mechanical and Manufacturing Engineering, University of New South Wales, Sydney, New South Wales 2052, Australia.
Implantable systems with chronic stability, high sensing performance, and extensive spatial-temporal resolution are a growing focus for monitoring and treating several diseases such as epilepsy, Parkinson's disease, chronic pain, and cardiac arrhythmias. These systems demand exceptional bendability, scalable size, durable electrode materials, and well-encapsulated metal interconnects. However, existing chronic implantable bioelectronic systems largely rely on materials prone to corrosion in biofluids, such as silicon nanomembranes or metals.
View Article and Find Full Text PDFCalcif Tissue Int
January 2025
Orthopaedic Research Laboratory, Department of Orthopedic Surgery and Traumatology, Odense University Hospital & Department of Clinical Research, University of Southern Denmark, V18-812B-1, Etage 1, Bygning 45.4, Nyt Sund, SDU Campus 5230, Odense, Denmark.
There is an increasing demand for a suitable bone substitute to replace current clinical gold standard autografts or allografts. Majority of previous studies have focused on the early effects of substitutes on bone formation, while information on their long-term efficacies remains limited. This study investigated the efficacies of natural hydroxyapatite (nHA) derived from oyster shells and synthetic hydroxyapatite mixed with collagen (COL/HA) or chitosan (CS/HA) on bone regeneration and implant fixation in sheep.
View Article and Find Full Text PDFTunis Med
December 2024
Department of trauma and orthopedics surgery. Mouwasat Hospital Dammam, 32263 Eastern Province Kingdom of Saudi Arabia.
Unlabelled: Introduction-Aim: Full-polyethylene glenosphere reverse shoulder arthroplasty (FP-RSA) known commonly as inverted-bearing concept utilizes a polyethylene glenosphere and metallic humeral liner. The aim was to assess mid-term outcomes of FP-RSA focusing on the incidence of scapular notching.
Methods: A retrospective analysis of 24 consecutive primary FP-RSA performed between 2017 and 2020 was conducted.
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!