Background: Pacemakers (PMs) and implantable cardioverter-defibrillators (ICDs) have reduced mortality and improved the quality of life of cardiac patients. However, the high cost of these devices prevents their large-scale incorporation, particularly in low-income countries, where reusing explanted PMs/ICDs has become an alternative.

Methods: A systematic review and meta-analysis were conducted of studies that compare infection rates, device-related deaths, malfunction and premature battery depletion in patients with reused PM and ICD implants and those with new devices. The quality of the body of evidence was assessed using the Grading of Recommendations Assessment, Development and Evaluation (GRADE) framework.

Results: The meta-analysis demonstrated no significant intergroup differences in infection rates (OR 0.98; 95% CI 0.60-1.60), device malfunction (OR 1.58; 95% CI 0.56-4.48) or premature battery depletion (OR 1.96; 95% CI 0.81-4.72) and no device-related deaths. Based on GRADE assessment, confidence in estimates for the outcomes infection rate and device-related death was rated as moderate.

Conclusion: The results of this analysis enabled us to conclude that PMs and ICDs can be safely and effectively reused. As such, every effort should be made to overcome regulatory, technical and ethical barriers to ensure implantation.

Download full-text PDF

Source
http://dx.doi.org/10.1080/17434440.2021.1927706DOI Listing

Publication Analysis

Top Keywords

implantable cardioverter-defibrillators
8
systematic review
8
review meta-analysis
8
body evidence
8
infection rates
8
device-related deaths
8
premature battery
8
battery depletion
8
reuse pacemakers
4
pacemakers implantable
4

Similar Publications

Pocket hematoma is a common and serious complication following cardiac implantable electronic device (CIED) implantation, contributing to significant morbidity and mortality. This study aimed to evaluate the efficacy of a novel pocket compression device in reducing pocket hematoma occurrence. We enrolled 242 patients undergoing CIED implantation, randomly assigning them to receive either the novel compression vest with a pressure cuff or conventional sandbag compression.

View Article and Find Full Text PDF

Since its first pathological description over 65 years ago, hypertrophic cardiomyopathy (HCM), with a worldwide prevalence of 1:500, has emerged as the most common genetically determined cardiac disease. Diagnostic work-up has dramatically improved over the last decades, from clinical suspicion and abnormal electrocardiographic findings to hemodynamic studies, echocardiography, contrast-enhanced cardiac magnetic resonance, and genetic testing. The implementation of screening programs and the use of implantable cardioverter defibrillators (ICDs) for high-risk individuals have notably reduced arrhythmic sudden deaths, altering the disease's mortality profile.

View Article and Find Full Text PDF

Background: The benefit of implantable cardioverter-defibrillators (ICD) and cardiovascular resynchronization therapy (CRT-D) in patients supported with a HeartMate 3 left ventricular assist device (LVAD) remains uncertain.

Methods: An analysis of the MOMENTUM 3 randomized clinical trial and the first 1000 patients in the Continued Access Protocol trial. Patients were divided into three groups based on the presence of ICD and/or CRT-D: No device (n=153, 11%), ICD only (n=699, 50.

View Article and Find Full Text PDF

Background: Chagas cardiomyopathy (CCM) is a significant cause of ventricular arrhythmias and sudden cardiac death (SCD). Although, implantable cardiac defibrillators (ICD) have been used for all forms of non-ischemic cardiomyopathy (NICM), studies on ICD efficacy in CCM are scarce.

Objective: The present study aims to compare the long-term outcomes, mortality rates, and the occurrence of tachycardia therapies after ICD implantation in patients with CCM and NICM.

View Article and Find Full Text PDF

Background: Patients with a left ventricular ejection fraction ≤ 35% are at increased risk of sudden cardiac death (SCD) within the first months after a myocardial infarction (MI). The wearable cardioverter defibrillator (WCD) is an established, safe and effective solution which can protect patients from SCD during the first months after an MI, when the risk of SCD is at its peak. This study aimed to evaluate the cost-effectiveness of WCD combined with guideline-directed medical therapy (GDMT) compared to GDMT alone, after MI in the English National Health Service (NHS).

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!