Purpose: Selective internal radiation therapy (SIRT) is a treatment for various kinds of liver tumours by injecting Y bearing microspheres into the liver vessels. To perform meaningful post-treatment dosimetry, quantitative imaging is performed.
Methods: This work uses a Monte-Carlo based reconstruction software with scatter and attenuation correction and collimator modelling that allows the quantification of Y bremsstrahlung SPECT/CT data. A dataset comprising 17 patients and measurements on a Jaszczak phantom, a NEMA IEC Body phantom and an anthropomorphic liver phantom are analysed and activities and dose values are acquired. These measured values are compared with applied activities and pre-treatment calculations, allowing to assess the quality of the SPECT reconstruction. A detailed uncertainty budget is presented, including uncertainties of the dose calibrator, the count rate, non-included interactions and other factors.
Results: The applied method is validated by finding measurements repeatable within the given uncertainty, and it is shown the influence of various parameters on the reconstruction process is negligible. Furthermore, activities and doses measured in the phantoms show good agreement with calculated values, if they are corrected for partial volume effects.
Conclusions: The strict observation of metrological requirements and the creation of an uncertainty budget increase the reliability and traceability of this novel approach to Y dosimetry. It gives an example of successful voxel-based dosimetry based on quantitative Y SPECT/CT image data.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.zemedi.2021.03.005 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!