Voltage waveform breakdown is characteristic of barotraumatic shock-wave generation during electrical catheter ablation of cardiac arrhythmias. The purpose of this investigation was to avoid barotrauma by defining, in vitro, the limits of pulse amplitude and pulse width for rectangular constant-current pulses that do not result in voltage breakdown and subsequently to determine what pulsing frequency is safe for use when high-energy trains of pulses are used. Electric pulses were delivered with a variable waveform modulator with a wide dynamic range and bandwidth capable of delivering pulses of 30-10,000-mu sec duration with amplitudes of up to 25 A. Cathodal pulses were delivered to a 6F catheter immersed in fresh anticoagulated bovine blood warmed to 37 degrees C to stimulate the milieu of a catheter in the chambers of the human heart. The maximum pulse amplitude that could be delivered without incurring voltage waveform breakdown varied inversely with pulse duration. Pulses of 30 mu sec broke down at currents above 24 A (2,500 V). Pulses of 10,000-mu sec duration broke down at 1 A (250 V). The maximum safely delivered energy for a single pulse was 2.5 J for pulses of 80-120 mu sec. Peak power for single pulses was maximum at 50-55 kW with 30-50-mu sec pulses. Charge delivery for single pulses was maximized at 9 mC with long, 10,000-mu sec duration pulses. To deliver an electrical pulse with energy significantly greater than 2.5 J without incurring voltage breakdown, trains of pulses were delivered where each pulse in the train had previously been shown to be free of voltage breakdown.(ABSTRACT TRUNCATED AT 250 WORDS)

Download full-text PDF

Source
http://dx.doi.org/10.1161/01.res.63.2.409DOI Listing

Publication Analysis

Top Keywords

pulses
13
voltage breakdown
12
pulses delivered
12
pulse
8
pulse width
8
shock-wave generation
8
voltage waveform
8
waveform breakdown
8
pulse amplitude
8
trains pulses
8

Similar Publications

Objective/AimIn 2009, dronedarone was approved by the United States Food and Drug Administration based on results from the ATHENA trial (NCT00174785), which showed significant reduction of cardiovascular (CV) hospitalization and death in patients with atrial fibrillation (AF) randomized to dronedarone versus placebo. In 2020, a retrospective study by Goehring et al. showed CV hospitalizations and deaths were lower in clinical practice following initiation of dronedarone compared to other antiarrhythmic drugs (AADs) in patients with AF and atrial flutter.

View Article and Find Full Text PDF

Bioinspired Conductivity-Enhanced, Self-Healing, and Renewable Silk Fibroin Hydrogel for Wearable Sensors with High Sensitivity.

ACS Appl Mater Interfaces

January 2025

Hebei Provincial Key Laboratory of Photoelectric Control on Surface and Interface, and College of Science, Hebei University of Science and Technology, Yuxiang Road 26, Shijiazhuang 050080, PR China.

The development of silk fibroin-based hydrogels with excellent biocompatibility, aqueous processability, and facile controllability in structure is indeed an exciting advancement for biological research and strain sensor applications. However, silk fibroin-based hydrogel strain sensors that combine high conductivity, high stretchability, reusability, and high selectivity are still desired. Herein, we report a simple method for preparing double-network hydrogels including silk fibroin and poly(acrylic acid) sodium-polyacrylate (PAA-PAAS) networks.

View Article and Find Full Text PDF

Background: The therapeutic armamentarium for heart failure with preserved ejection fraction (HFpEF) remains notably constrained. A factor contributing to this problem could be the scarcity of in vitro models for HFpEF, which hinders progress in developing new therapeutic strategies. Here, we aimed at developing a novel, comorbidity-inspired, human, in vitro model for HFpEF.

View Article and Find Full Text PDF

Background: Cardioneuroablation (CNA) is a new approach to treat reflex syncope and functional bradyarrhytmias caused by autonomic imbalance. We report our experience using CNA.

Method: From September 2022 to July 2023, we took care of 21 patients (mean age 42 ± 21 years; 62% male) affected by reflex syncope or functional bradyarrhythmias.

View Article and Find Full Text PDF

Background: Awake prone positioning (APP) can reportedly reduce the need for intubation and help improve prognosis of patients with acute hypoxemic respiratory failure (AHRF) infected with COVID-19. However, its physiological mechanism remains unclear. In this study, we evaluated the effect of APP on lung ventilation in patients with moderate-to-severe AHRF to better understand the effects on ventilation distribution and to prevent intubation in non-intubated patients.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!