Importance: Scaphoid fractures are the most common carpal fracture, but as many as 20% are not visible (ie, occult) in the initial injury radiograph; untreated scaphoid fractures can lead to degenerative wrist arthritis and debilitating pain, detrimentally affecting productivity and quality of life. Occult scaphoid fractures are among the primary causes of scaphoid nonunions, secondary to delayed diagnosis.
Objective: To develop and validate a deep convolutional neural network (DCNN) that can reliably detect both apparent and occult scaphoid fractures from radiographic images.
Design, Setting, And Participants: This diagnostic study used a radiographic data set compiled for all patients presenting to Chang Gung Memorial Hospital (Taipei, Taiwan) and Michigan Medicine (Ann Arbor) with possible scaphoid fractures between January 2001 and December 2019. This group was randomly split into training, validation, and test data sets. The images were passed through a detection model to crop around the scaphoid and were then used to train a DCNN model based on the EfficientNetB3 architecture to classify apparent and occult scaphoid fractures. Data analysis was conducted from January to October 2020.
Exposures: A DCNN trained to discriminate radiographs with normal and fractured scaphoids.
Main Outcomes And Measures: Area under the receiver operating characteristic curve (AUROC), sensitivity, and specificity. Fracture localization was assessed using gradient-weighted class activation mapping.
Results: Of the 11 838 included radiographs (4917 [41.5%] with scaphoid fracture; 6921 [58.5%] without scaphoid fracture), 8356 (70.6%) were used for training, 1177 (9.9%) for validation, and 2305 (19.5%) for testing. In the testing test, the first DCNN achieved an overall sensitivity and specificity of 87.1% (95% CI, 84.8%-89.2%) and 92.1% (95% CI, 90.6%-93.5%), respectively, with an AUROC of 0.955 in distinguishing scaphoid fractures from scaphoids without fracture. Gradient-weighted class activation mapping closely corresponded to visible fracture sites. The second DCNN achieved an overall sensitivity of 79.0% (95% CI, 70.6%-71.6%) and specificity of 71.6% (95% CI, 69.0%-74.1%) with an AUROC of 0.810 when examining negative cases from the first model. Two-stage examination identified 20 of 22 cases (90.9%) of occult fracture.
Conclusions And Relevance: In this study, DCNN models were trained to identify scaphoid fractures. This suggests that such models may be able to assist with radiographic detection of occult scaphoid fractures that are not visible to human observers and to reliably detect fractures of other small bones.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC8103226 | PMC |
http://dx.doi.org/10.1001/jamanetworkopen.2021.6096 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!