The irregular shape and uneven tissue thickness of excised breast specimens makes radiographic evaluation difficult, especially when calcifications are not present. Xeroradiographs before and after compression of 20 separate excised breast specimens were compared, and 17 of the same specimens were compared after compression combined with immersion in water. Specimen compression improved visibility of the lesion on average in 88% of cases, and visibility was equal in 12%. Combined compression/immersion further improved visibility of the lesion on average in 37% of cases. More significantly, evaluation of the compressed specimen led to a change in interpretation of the radiographs in 45% of cases. Compression of the specimen in specimen radiography is recommended in all cases in which pre-biopsy localization is performed.

Download full-text PDF

Source
http://dx.doi.org/10.1148/radiology.168.2.3393660DOI Listing

Publication Analysis

Top Keywords

specimen radiography
8
excised breast
8
breast specimens
8
specimens compared
8
improved visibility
8
visibility lesion
8
lesion average
8
compression
5
breast specimen
4
radiography evaluation
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!