The LAmbre device is a novel system designed for left atrial appendage closure (LAAC). First registries showed a high rate of device implantation success. However, few mid-term results are available. We present our 1- and 12-month follow-up results for this device. This prospective, single-center registry included consecutive patients with nonvalvular atrial fibrillation who underwent LAAC with the LAmbre device. Transesophageal echocardiography (TEE) was performed at 1-month follow-up. In total, 55 patients were included. The population was elderly (75 ± 9.4 years), with a high proportion of comorbidities. The mean CHADS-VASc and HAS-BLED scores were 4.6 ± 1.6 and 3.9 ± 1.0, respectively. Previous history of a major bleeding event was present in 37 patients (67.3%). Procedural success was achieved in 54 patients (98.2%). Device success was achieved in 100% of patients in whom device implantation was attempted (54 patients). Major in-hospital device-related complications included mortality of one patient (1.8%) and pericardial tamponade in two patients (3.6%); the incidence of stroke was 0%. No thrombus or significant leaks (≥5 mm) were observed on 1-month TEE. At 12 months, adverse events were overall death (1.8%), transient ischemic attack/ischemic stroke (1.8%), and major bleeding events (Bleeding Academic Research Consortium (BARC) 3a and 3c; 11%). In this high-risk population, the LAmbre device seems to be a safe and effective option for LAAC with a remarkable mid-term performance.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC8036586PMC
http://dx.doi.org/10.3390/jcm10071421DOI Listing

Publication Analysis

Top Keywords

lambre device
12
left atrial
8
atrial appendage
8
appendage closure
8
device
8
mid-term performance
8
device implantation
8
major bleeding
8
success achieved
8
patients
7

Similar Publications

Intracardiac vs transesophageal echocardiography for left atrial appendage occlusion: An updated systematic review and meta-analysis.

Heart Rhythm

August 2024

Richard A. and Susan F. Smith Center for Outcomes Research, Beth Israel Deaconess Medical Center and Harvard Medical School, Boston, Massachusetts; Division of Cardiology, Beth Israel Deaconess Medical Center, Harvard Medical School, Boston, Massachusetts. Electronic address:

Background: Multiple studies continue to evaluate the use of intracardiac echocardiography (ICE) and transesophageal echocardiography (TEE) for guiding left atrial appendage occlusion (LAAO).

Objective: The purpose of this study was to conduct an updated meta-analysis comparing the effectiveness and safety outcomes of both imaging modalities.

Methods: PubMed, Cochrane, and Embase were searched for studies comparing ICE vs TEE to guide LAAO.

View Article and Find Full Text PDF

Clinical outcome and intraprocedural characteristics of left atrial appendage occlusion: a comparison between single-occlusive plug-type and dual-occlusive disc-type devices.

Front Cardiovasc Med

July 2024

Deutsches Herzzentrum der Charité, Department of Cardiology, Angiology and Intensive Care Medicine, Campus Virchow Klinikum, Berlin, Germany.

Article Synopsis
  • Percutaneous interventional left atrial appendage occlusion (LAAO) is a proven method for preventing strokes in patients with atrial fibrillation (AF) and has been evaluated in a study involving 149 patients at a Berlin hospital from 2016 to 2022.
  • The study compared two types of devices used in LAAO: single-occlusive plug-type (SOPT) and dual-occlusive disc-type (DODT), noting that while DODT procedures took longer, both had successful implantation with low rates of complications and no hospital deaths.
  • After six months, findings indicated that SOPT had a higher rate of device-related thrombus formation compared
View Article and Find Full Text PDF

Background: The goal of this study was to compare the procedural safety and long-term outcome associated with a combined catheter ablation and left atrial appendage occlusion (LAAO) procedure utilizing intracardiac echocardiography (ICE) guidance versus transesophageal echocardiography (TEE) guidance. The study focuses on implementing LAmbre and Watchman devices in patients diagnosed with nonvalvular atrial fibrillation (AF).

Methods: A total of 363 patients diagnosed with nonvalvular AF and who underwent a combined procedure were prospectively enrolled between November 2017 and May 2022.

View Article and Find Full Text PDF

Patients with previous interatrial shunt device (IASD) implantation may face greater challenges during future left atrial interventional procedures. Herein, we report the first case of left atrial appendage closure (LAAC) in a patient with previous IASD implantation. The patient successfully underwent LAAC using the LAmbre device without complications.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?