Purpose: This numerical study examined the efficacy and challenges of using resonance frequency analysis to identify the stability of implants placed in mandibles. The study also examined the feasibility of using angular stiffness as an alternative index to quantify dental implant stability in mandibles.
Materials And Methods: A finite element model consisting of a mandible, an implant, an abutment, and a bonding layer (between the implant and the mandible) was created in commercially available software ANSYS. The level of osseointegration was modeled by varying the stiffness of the bonding layer. Three sets of boundary conditions were imposed on the mandible: fixed, rotationally free, and rotationally restrained. Three implant locations were studied: central, premolar, and molar positions. An alternative abutment mimicking SmartPeg and eight different implant lengths were also included. A modal analysis and a static analysis were conducted to calculate resonance frequencies and angular stiffness, respectively.
Results: Two types of vibration modes were found. One was jawbone modes, for which the mandible deformed significantly but not the bonding layer. Resonance frequencies of the jawbone modes were not sensitive to the level of osseointegration. The other was implant modes, for which the bonding layer deformed significantly but not the mandible. Among multiple implant modes obtained, only one was trackable as the level of osseointegration increased. The resonance frequency of the trackable implant mode was very sensitive to the implant location as well as boundary conditions, but not as much to the level of osseointegration. In contrast, angular stiffness was sensitive to the level of osseointegration but not as much to boundary conditions.
Conclusion: The efficacy of using resonance frequency analysis to quantify the stability of a dental implant is questionable. Its high sensitivity to implant locations and boundary conditions as well as its low sensitivity to the level of osseointegration cause huge uncertainties in correlating measured resonance frequencies to implant stability. Angular stiffness is a much more reliable indicator because of its high sensitivity to the level of osseointegration and low sensitivity to boundary conditions.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.11607/jomi.8579 | DOI Listing |
J Dent Sci
December 2024
Division for Globalization Initiative, Liaison Center for Innovative Dentistry, Tohoku University Graduate School of Dentistry, Sendai, Japan.
Background/purpose: Titanium dioxide nanotube (TNT) structures have been shown to enhance the early osseointegration of dental implants. Nevertheless, the optimal nanotube diameter for promoting osteogenesis remains unclear due to variations in cell types and manufacture of nanotubes. This study aimed to evaluate the differences in MC3T3-E1 and Saos-2 cells behavior on nanotubes of varying diameters.
View Article and Find Full Text PDFObjectives: The goal of this systematic review was to critically appraise the existing evidence evaluating osteoporosis' effects on dental implant osseointegration and survival rate.
Data Source: A search was conducted in two databases, PubMed/MEDLINE and Scopus, until October 2024, using the keywords 'osteoporosis,' 'osteopenia,' 'osseointegration,' and 'dental implants'. The inclusion criteria were clinical studies that evaluated the implant placement, complications, and osseointegration results in patients with osteoporosis; literature reviews and clinical studies addressing the outcome were considered; and articles written in English and published since 2000.
J Adv Prosthodont
December 2024
Department of Prosthodontics, Dental Research Institute, Dental and Life Sciences Institute, Education and Research Team for Life Science on Dentistry, School of Dentistry, Pusan National University, Yangsan, Republic of Korea.
Purpose: This pilot study investigated the effect of surface roughness on osseointegration by comparing two types of commercial SLA-treated dental implants with different surface roughness levels: moderately rough (S = 1 - 2 µm) and rough surfaces (S > 2 µm).
Materials And Methods: Two implant groups were studied: TS (rough surface) and ADD (moderately rough surface) groups. Surface characteristics were analyzed using optical profilometry and SEM.
J Clin Med
December 2024
Oral Surgery Department, MALO CLINIC, Avenida dos Combatentes, 43, Level 9, 1600-042 Lisboa, Portugal.
: In the last decades, dental implant surfaces have been evolving to increase success and implant survival rates. More studies evaluating outcomes with implants with ultra-hydrophilic multi-zone anodized surfaces are necessary. The aim of this study is to evaluate the short-term outcome of implants of conical connection with anodized ultra-hydrophilic surfaces for support of single teeth and partial rehabilitations.
View Article and Find Full Text PDFJ Contemp Dent Pract
September 2024
Department of Pediatrics Dentistry and Orthodontics, Faculty Odonto-Stomatology, Can Tho University of Medicine and Pharmacy, Can Tho City, Vietnam.
Aim: This study aimed to evaluate the impact of a combination of immediate implant placement with maxillary sinus augmentation (MSA) solely using platelet-rich fibrin (PRF) on guided bone regeneration.
Materials And Methods: An interventional before-after (pre-post) study design was used with 30 dental patients (≥18 years of age; 14 males and 16 females) with initial bone heights ranging between 4 and 6 mm. Following the general check-up and the creation of a study model, the planned implant location demonstrated an external right maxilla diameter of more than 5 mm, thereby validating the cone-beam computed tomography (CBCT) radiograph.