Verification of field match lines in whole breast radiation therapy using Cherenkov imaging.

Radiother Oncol

Norris Cotton Cancer Center at Dartmouth Hitchcock Medical Center, Lebanon, United States; Geisel School of Medicine, Dartmouth College, Hanover, United States. Electronic address:

Published: July 2021

Purpose: In mono-isocentric radiation therapy treatment plans designed to treat the whole breast and supraclavicular lymph nodes, the fields meet at isocenter, forming the match line. Insufficient coverage at the match line can lead to recurrence, and overlap over weeks of treatment can lead to increased risk of healthy tissue toxicity. Cherenkov imaging was used to assess the accuracy of delivery at the match line and identify potential incidents during patient treatments.

Methods And Materials: A controlled calibration was constructed from the deconvolved Cherenkov images from the delivery of a modified patient treatment plan to an anthropomorphic phantom with introduced separation and overlap. The trend from this calibration was then used to evaluate the field match line for accuracy and inter-fraction consistency for two patients.

Results: The intersection point between matching field profiles was directly correlated to the distance (gap/overlap) between the fields (anthropomorphic phantom R = 0.994 "breath hold" and R = 0.990 "free breathing"). The profile intersection points from two patients' imaging sessions yielded an average of +1.40 mm offset (overlap) and -1.32 mm offset (gap), thereby introducing roughly a 25.0% over-dose and a -23.6% under-dose (R = 0.994).

Conclusions: This study shows that field match regions can be detected and quantified by taking deconvolved Cherenkov images and using their product image to create steep intensity gradients, causing match lines to stand out. These regions can then be quantitatively translated into a dose consequence. This approach offers a high sensitivity detection method which can quantify match line variability and errors in vivo.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC8686843PMC
http://dx.doi.org/10.1016/j.radonc.2021.04.013DOI Listing

Publication Analysis

Top Keywords

field match
12
match
8
match lines
8
radiation therapy
8
cherenkov imaging
8
deconvolved cherenkov
8
cherenkov images
8
anthropomorphic phantom
8
verification field
4
lines breast
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!