Purpose: Prostate cancer is the second most prevalent cancer in US men, with about 192,000 new cases and 33,000 deaths predicted for 2020. With only a 31% 5-year survival rate for patients with an initial diagnosis of stage-four prostate cancer, the necessity for early screening and diagnosis is clear. In this paper, we present navigation accuracy results for Promaxo's MR system intended to be used in a physician's office for image-guided transperineal prostate biopsy.

Methods: The office-based low-field MR system was used to acquire images of prostate phantoms with needles inserted through a transperineal template. Coordinates of the estimated sample core locations in the office-based MR system were compared to ground truth needle coordinates identified in a 1.5T external reference scan. The error was measured as the distance between the planned target and the ground truth core center and as the shortest perpendicular distance between the planned target and the ground truth trajectory of the whole core.

Results: The average error between the planned target and the ground truth core center was 2.57 ± 1.02 mm, [1.93-3.21] 95% CI. The average error between the planned target to the actual core segment was 2.05 ± 1.24 mm, [1.53-2.56] 95% CI.

Conclusion: The average navigation errors were below the clinically significant threshold of 5 mm. The initial phantom results demonstrate the feasibility of the office-based system for prostate biopsy.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC8134310PMC
http://dx.doi.org/10.1007/s11548-021-02364-7DOI Listing

Publication Analysis

Top Keywords

ground truth
16
planned target
16
target ground
12
initial phantom
8
system prostate
8
prostate biopsy
8
prostate cancer
8
office-based system
8
distance planned
8
truth core
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!