Background: Left bundle branch pacing (LBBP) is a novel conduction system pacing modality, but pacing lead deployment remains challenging.
Objectives: This study aimed to evaluate the feasibility of visualization-enhanced lead deployment for LBBP implantation and to assess LBBP characteristics on the basis of lead tip location.
Methods: Successful LBBP with a well-defined lead tip location by visualization of the tricuspid value annulus in 20 patients was retrospectively analyzed to develop an image-guided technique to identify the LBBP target site. This technique was then prospectively tested in 60 patients who were randomized into 2 groups, one using the standard approach (the standard group) and the other using the image-guided technique (the visualization group). The procedural details, electrophysiological characteristics, and short-term follow-up were compared between groups.
Results: LBBP was successfully achieved in 28 patients in the standard group and in 29 in the visualization group. The procedural and fluoroscopic durations in the visualization group (66.76 ± 14.62 and 7.83 ± 2.05 minutes) were significantly shorter than those in the standard group (85.46 ± 20.19 and 11.11 ± 3.51 minutes) (P < .01). The number of lead deployment attempts in the visualization group was lower than that in the standard group (2.03 ± 1.18 vs 2.96 ± 1.17; P < .01), and the proportion of left bundle branch potential recorded was higher (79.3% vs 46.4%; P = .01).
Conclusion: Using a visualization technique, the procedural and fluoroscopic durations for LBBP implantation were significantly shortened with fewer lead repositioning attempts.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.hrthm.2021.04.015 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!