Efficacy of breast shielding during head computed tomography examination.

Radiol Oncol

University of Ljubljana, Faculty of Health Sciences, Medical Imaging and Radiotherapy Department, Ljubljana, Slovenia.

Published: August 2020

Background: Female breasts are exposed to scattered radiation regardless of not being included in the primary field during head CT. This study aimed to investigate whether the use of lead shielding is beneficial in dose reduction to the breasts during head CT.

Patients And Methods: The study was performed in two different hospitals on two different CT units and included 120 patients. Half of the measurements (n = 60) was conducted without the use of lead shielding and the other half (n = 60) with the use of lead shielding of 0.5 mm equivalent thickness.

Results: Significant skin dose reduction to the breasts during head CT in both hospitals with the use of lead shielding was discovered; 81% (338.2 ± 43.7 μGy to 64.3 ± 18.8 μGy) in Hospital A and 74% (from 253.1 ± 35.1 μGy to 65.3 ± 16.9 μGy) in Hospital B.

Conclusions: Considering the assumed carcinogenic effect of low doses of radiation, high frequency of the head CT scans and the significant reduction of radiation doses to the highly radiosensitive breasts, the use of lead shielding is highly recommendable.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC7877269PMC
http://dx.doi.org/10.2478/raon-2020-0044DOI Listing

Publication Analysis

Top Keywords

lead shielding
20
dose reduction
8
reduction breasts
8
breasts head
8
μgy hospital
8
shielding
6
head
5
lead
5
efficacy breast
4
breast shielding
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!