Objective: Schistosomiasis japonica is an important helminthic disease in Asia. Sensitive and accurate diagnostic tools are indispensable for clinical diagnosis, screening infection and monitoring its control. In this study, we developed an immunochromatographic test (Sj-ICT) to detect anti-Schistosoma japonicum immunoglobulin G antibodies in human sera.

Methods: Somatic extract from adult S. japonicum was used as an antigen. The Sj-ICT was developed and optimized as a point-of-care test. All 214 human serum samples were evaluated for diagnostic usefulness and comparison with an enzyme-linked immunosorbent assay (ELISA).

Results: The diagnostic sensitivity, specificity, positive and negative predictive values, and accuracy of the Sj-ICT were 90.8%, 87.9%, 86.4%, 91.9% and 89.3%, respectively. For ELISA the values were respectively 91.8%, 87.9%, 86.5%, 92.7% and 89.7%. The concordance between both methods was 86.4 % (Cohen's kappa value = 0.729).

Conclusions: The immunochromatographic test kit developed can support clinical diagnosis and large-scale surveys in endemic areas without requiring additional facilities or ancillary supplies.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.ijid.2021.04.039DOI Listing

Publication Analysis

Top Keywords

point-of-care test
8
enzyme-linked immunosorbent
8
immunosorbent assay
8
immunoglobulin antibodies
8
schistosomiasis japonica
8
clinical diagnosis
8
immunochromatographic test
8
comparison point-of-care
4
test
4
test enzyme-linked
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!