Objective: Preoperative digital templating is a standard procedure in total hip arthroplasty. Deviations between template size and final implant size may result from inaccurate calibration, templating as well as intraoperative decisions. So far, the explicit effect of calibration errors on templating has not been addressed adequately.

Materials And Methods: A mathematical simulation of calibration errors up to ± 24% was applied to the templating of acetabular cups (38 to 72 mm diameter). The effect of calibration errors on template component size as deviation from optimal size was calculated.

Results: The relationship between calibration error and component size deviation is inverse and linear. Calibration errors have a more pronounced effect on larger component sizes. Calibration errors of 2-6% result in templating errors of up to two component sizes. Common errors of up to 12% may result in templating errors of 3-4 sizes for common implant sizes. A tabular matrix visualizes the effect.

Conclusion: Calibration errors play a significant role in component size selection during digital templating. Orthopedic surgeons should be aware of this effect and try to identify and address this source of error.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC8166734PMC
http://dx.doi.org/10.1007/s11548-021-02367-4DOI Listing

Publication Analysis

Top Keywords

calibration errors
24
component size
16
digital templating
12
calibration
9
errors
9
calibration error
8
templating
8
total hip
8
hip arthroplasty
8
size deviation
8

Similar Publications

Significance: Laparoscopic surgery presents challenges in localizing oncological margins due to poor contrast between healthy and malignant tissues. Optical properties can uniquely identify various tissue types and disease states with high sensitivity and specificity, making it a promising tool for surgical guidance. Although spatial frequency domain imaging (SFDI) effectively measures quantitative optical properties, its deployment in laparoscopy is challenging due to the constrained imaging environment.

View Article and Find Full Text PDF

Background: Wheat-maize cropping systems in semi-arid regions are expected to be affected by climate change in the future, which is alarming for global food security, environmental sustainability and socioeconomic development. Therefore, management practices like optimized plant geometry and fertilization need to be explored to counter these expected threats. To do this, the APSIM model was calibrated using 5-year data (from 2017/2018 to 2022) regarding yield, biomass, plant height, emergence, anthesis and crop maturity of wheat and maize from farmer fields.

View Article and Find Full Text PDF

Background: Histotripsy is a non-invasive, non-ionizing, non-thermal focused ultrasound technique. High amplitude short acoustic pulses converge to create high negative pressures that cavitate endogenous gas into a bubble cloud leading to mechanical tissue destruction. In the United States, histotripsy is approved to treat liver tumors under diagnostic ultrasound guidance but in initial clinical cases, some areas of the liver have not been treated due to bone or gas obstructing the acoustic window for targeting.

View Article and Find Full Text PDF

Background: The use of iodinated contrast-enhancing agents in computed tomography (CT) improves the visualization of relevant structures for radiotherapy treatment planning (RTP). However, it can lead to dose calculation errors by incorrectly converting a CT number to electron density.

Purpose: This study aimed to propose an algorithm for deriving virtual non-contrast (VNC) electron density from dual-energy CT (DECT) data.

View Article and Find Full Text PDF

Count-rate management in I SPECT/CT calibration.

EJNMMI Phys

January 2025

Department of Medical Radiation Physics and Nuclear Medicine, Karolinska University Hospital, Solna, Sweden.

Background: System calibration is essential for accurate SPECT/CT dosimetry. However, count losses due to dead time and pulse pileup may cause calibration errors, in particular for I, where high count rates may be encountered. Calibration at low count rates should also be avoided to minimise detrimental effects from e.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!