Severity: Warning
Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 176
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 1034
Function: getPubMedXML
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3152
Function: GetPubMedArticleOutput_2016
File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 316
Function: require_once
Background: Bi-modular stems were introduced in primary total hip arthroplasty (THA) to enable better control of the femoral offset, leg length, and hip stability. Despite numerous reports on modular femoral neck fractures, some designs are still marketed worldwide. While the risk factors for the sudden failure are multifactorial and mostly known, the timing of this new THA complication is not predictable by any means.
Case Presentation: In this report, the literature regarding one of the most popular bi-modular stems with specific neck-stem coupling (oval Morse taper) is reviewed and illustrated with a case of bilateral modular neck fracture in a patient with idiopathic aseptic necrosis of femoral heads treated with primary bi-modular THA. Because of bilateral modular femoral neck fracture, which occurred 3 years on the left side and 20 years after implantation on the right side, the patient required a total of 6 revisions and 208 days of hospitalized care.
Conclusion: To our knowledge, this is the first report of bilateral modular neck fracture in a single patient. Even though the same surgeon performed both operations and used the same neck length and orientation, fractures occurred with a 17-year time difference after implantation. This shows that we cannot predict with certainty when a fracture might occur. Orthopaedic surgeons should use bi-modular stem designs for primary THA very cautiously.