Gimme' the Loop: Modifications to the Squid-Capture Technique for Laser-Fenestrated Thoracic Endografting.

J Endovasc Ther

Department of Vascular and Endovascular Surgery, Medical Faculty and University Hospital Düsseldorf, Heinrich-Heine-University, Düsseldorf, Germany.

Published: August 2021

Purpose: Laser-fenestrated thoracic endovascular aortic repair (LfTEVAR) in the aortic arch with covering of the left subclavian artery (LSA) orifice is challenging. To optimize fenestration, the so-called squid-capture technique has been introduced. We present here a modification to the technique that may help improve time-efficiency and safety.

Technique:: During the originally proposed squid-capture maneuver, the stent-graft is deployed in a preset snare wire loop, which is used to pull the stent graft toward the penetration device during in-situ fenestration. In preparation, the guidewire needs to be passed through the loop inside the aortic arch, which can be difficult and may predispose for embolic events. We propose here the creation of a "guidewire-through-snare-loop" configuration outside the body, which can then be reliably transferred into the aortic arch. The modified technique was successfully applied in a patient undergoing LfTEVAR for penetrating aortic ulcers.

Conclusion: The proposed modification may help facilitate the squid-capture technique for LfTEVAR while saving time and resources. Given that LfTEVAR is becoming more frequently used, it is important to ensure technical success and safety of the procedure.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC8276338PMC
http://dx.doi.org/10.1177/15266028211007475DOI Listing

Publication Analysis

Top Keywords

squid-capture technique
12
aortic arch
12
laser-fenestrated thoracic
8
technique
5
aortic
5
gimme' loop
4
loop modifications
4
squid-capture
4
modifications squid-capture
4
technique laser-fenestrated
4

Similar Publications

The effect of in situ laser fenestration for total endovascular arch repair in redo aortic dissection.

Vascular

December 2022

Department of Cardiovascular Surgery, Second Affiliated Hospital, 89681Zhejiang University, Hangzhou, People's Republic of China.

Objective: Treatment of aortic arch pathologies in redo cases is technically challenging. In this study, we assessed early and mid-term outcomes of total endovascular arch repair combined with a new method of in situ laser fenestration.

Methods: Between January 2018 and March 2019, five patients with a history of cardiovascular surgery underwent in situ laser fenestration procedures using the "squid capture technique" for aortic arch pathologies with dissection.

View Article and Find Full Text PDF

Gimme' the Loop: Modifications to the Squid-Capture Technique for Laser-Fenestrated Thoracic Endografting.

J Endovasc Ther

August 2021

Department of Vascular and Endovascular Surgery, Medical Faculty and University Hospital Düsseldorf, Heinrich-Heine-University, Düsseldorf, Germany.

Purpose: Laser-fenestrated thoracic endovascular aortic repair (LfTEVAR) in the aortic arch with covering of the left subclavian artery (LSA) orifice is challenging. To optimize fenestration, the so-called squid-capture technique has been introduced. We present here a modification to the technique that may help improve time-efficiency and safety.

View Article and Find Full Text PDF

We present a case of an 85-year-old woman with bilateral limb-threatening ischemia caused by acute-on-chronic occlusion of the infrarenal aorta. The patient once underwent endovascular recanalization using nitinol and stainless-steel bare-metal stent implantation; however, the stainless-steel stent collapsed 3 months later. In the second endovascular therapy, "Squid-Capture" modified in situ stent-graft fenestration technique followed by stent-in-stent implantation with stent graft and bare-metal stent was successfully applied, and it can be regarded as a promising treatment option for the repair of abdominal aortic occlusive disease in some limited anatomical conditions.

View Article and Find Full Text PDF

[Debranch Thoracic Endovascular Aortic Therapy for Extending Aneurysms].

Kyobu Geka

July 2016

Department of Cardiovascular Surgery, Oita University, Oita, Japan.

To apply endovascular aortic repair for arch or thoracoabdominal pathology, it is essential to reconstruct the branches originating in the treatment area. In cases that a stentgraft has to reach ascending aorta we perform "in situ fenestration with squid capture technique".During the procedure cerebral circulation is maintained by percutaneous cardiopulmonary bypass.

View Article and Find Full Text PDF

An 83-year-old female was found to have an fusiform aneurysm in the aortic arch. She was deemed to be a high surgical risk; therefore, endovascular stent-graft placement followed by revascularization of the brachiocephalic trunk using in situ stent-graft fenestration was considered. However, the safe application of fenestration was deemed difficult due to the tortuosity of the brachiocephalic artery.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!