Objective: The aim of this study was to provide early-term evaluation, safety, and efficacy of the novel CGuard (Inspire MD, Tel Aviv, Israel) micromesh self-expanding stent with embolic protection system (EPS) during carotid artery stenting (CAS) procedure.

Materials And Methods: All patients who underwent CAS with CGuard carotid stent system from January 2018 to September 2020 in a tertiary center were prospectively evaluated. Primary endpoints included technical success and perioperative neurological events (0-48 hours). Secondary endpoint was the rate of neurologic, cardiac events, and death (major adverse event or MAE) at 30 days. Patency of CGuard, peak systolic velocity (PSV), and end diastolic velocity (EDV) were evaluated at 30 days with duplex ultrasound.

Results: A total of 103 patients underwent CAS procedure. Forty patients were symptomatic, and 63 patients were asymptomatic with stenosis greater than 70%. Technical success was 100%. Local anesthesia was applied in 94 patients (93%). Embolic protection devices were used in 6 patients (5.8%). Access site complications were 1.9%. Mean internal carotid artery stenosis diameter reduced from 79.47% to 14.68%. Two patients had transient ischemic attack (1.9%) and 1 patient experienced a cerebral hyperperfusion syndrome (CHS) perioperatively. External carotid artery (ECA) occlusion occurred in 1.9%. Four patients experienced jaw claudication (3.8%) postoperatively. Mean time of operation was 41 minutes. Mean duration of hospitalization was 3.1 days. The 30-day rate of MAE was 0%. CGuard patency was 100%, mean internal carotid PSV was reduced from 251.57 to 77.29 cm/s, and mean internal carotid EDV was reduced from 154.62 to 24.63 cm/s at 30 days.

Conclusion: Our study shows that CGuard stent with EPS is an effective and safe device for treatment of carotid artery stenosis with acceptable low perioperative neurologic events, even with low embolic protection device usage. Larger multicenter and randomized studies are necessary to confirm its long-term efficacy.

Download full-text PDF

Source
http://dx.doi.org/10.1177/15266028211007466DOI Listing

Publication Analysis

Top Keywords

carotid artery
20
embolic protection
12
internal carotid
12
patients
9
carotid
8
artery stenting
8
patients underwent
8
underwent cas
8
technical success
8
artery stenosis
8

Similar Publications

A 69-year-old man with chest pain was diagnosed with acute type B aortic dissection with the entry tear located at distal arch and a distal aortic arch aneurysm. Therefore, we performed debranching thoracic endovascular aortic repair 2 weeks after type B aortic dissection onset. First, the graft was anastomosed to bilateral axillary arteries.

View Article and Find Full Text PDF

Fibromuscular dysplasia (FMD) is a non-atherosclerotic, non-inflammatory vascular disease of medium-sized arteries that causes abnormal cellular growth in arterial walls and most commonly affects young to middle-aged women (20-50 years of age). While FMD often involves the renal arteries, it can affect any arterial bed. FMD has a characteristic angiographic appearance of a "string of beads.

View Article and Find Full Text PDF

Objective: To explore the anatomical and clinical factors that affect the radiographic exposure time in radial artery cerebral angiography and to establish a model.

Method: A total of 210 patients who underwent radial artery cerebral angiography at this center from September 2021 to May 2022 were selected, and their anatomical and clinical factors were analyzed to evaluate the correlation between these factors and the duration of radiographic exposure. A related neural network prediction model was established.

View Article and Find Full Text PDF

Purpose: To describe a case of short common trunk of the occipital artery (OA) and ascending pharyngeal artery (APA) arising from the internal carotid artery (ICA).

Methods: A 36-year-old woman with a history of surgical resection of a right lateral ventricular meningioma and atheromatous plaque of the right ICA underwent cranial magnetic resonance (MR) imaging and MR angiography of the head and neck region with a 3-Tesla scanner.

Results: MR angiography of the neck region showed a small atheromatous plaque at the origin of the right ICA and an anomalous artery arising from the posteromedial aspect of the right ICA at the distal end of the carotid bulb.

View Article and Find Full Text PDF

Programmable scanning diffuse speckle contrast imaging of cerebral blood flow.

Neurophotonics

January 2025

University of Kentucky, Department of Biomedical Engineering, Lexington, Kentucky, United States.

Significance: Cerebral blood flow (CBF) imaging is crucial for diagnosing cerebrovascular diseases. However, existing large neuroimaging techniques with high cost, low sampling rate, and poor mobility make them unsuitable for continuous and longitudinal CBF monitoring at the bedside.

Aim: We aimed to develop a low-cost, portable, programmable scanning diffuse speckle contrast imaging (PS-DSCI) technology for fast, high-density, and depth-sensitive imaging of CBF in rodents.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?