MONTE CARLO DOSE ASSESSMENT IN DENTAL CONE-BEAM COMPUTED TOMOGRAPHY.

Radiat Prot Dosimetry

School of Mechanical Engineering, Pusan National University, Busan 46241, Republic of Korea.

Published: May 2021

Most dental cone-beam computed tomography (CBCT) uses an x-ray beam field covering the maxillomandibular region and the width-truncated detector geometry. The spatial dose distribution in dental CBCT is analyzed in terms of local primary and remote secondary doses by using a list-mode analysis of x-ray interactions obtained from the Monte Carlo simulations. The patient-dose benefit due to the width-truncated detector geometry is also investigated for a wide range of detector offsets. The developed dose estimation agrees with the measurement in a relative error of 7.7%. The secondary dose outside of the irradiation field becomes larger with increasing tube voltage. The dose benefit with the width-truncated geometry linearly increases as the detector-offset width is decreased. Leaving the CT image quality out of the account, the MC results reveal that the operation of dental CBCT with a lower tube voltage and a smaller detector-offset width is beneficial to the patient dose.

Download full-text PDF

Source
http://dx.doi.org/10.1093/rpd/ncab039DOI Listing

Publication Analysis

Top Keywords

monte carlo
8
dental cone-beam
8
cone-beam computed
8
computed tomography
8
width-truncated detector
8
detector geometry
8
dental cbct
8
benefit width-truncated
8
tube voltage
8
detector-offset width
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!