Purpose: The in-room laser which is used for patient positioning in radiotherapy is generally projected on the radiation isocenter determined by the Winston-Lutz test and so on. In this study, a couch-mounted verification device was developed that could evaluate all in-room lasers' alignment including the horizontality and verticality at one time. The device has the function to perform the light/radiation field coincidence test and the Winston-Lutz test at the same time. The aim of this report was to introduce the verification procedure for two tests, using the newly developed software and device, and to present the tuning flow of the in-room laser. Moreover, the analysis accuracy of the developed software was evaluated in comparison with commercial software.
Methods: First, the light/radiation field was evaluated by using tungsten markers on the central surface of the device. Next, after aligning the long-carved lines on the front and sides of the device with the in-room lasers, the Winston-Lutz test was carried out by using the tungsten sphere in the center of the device. The acquired images were collectively analyzed using the developed software equipped with the reporting function. Additionally, the result of this Winston-Lutz test was compared with the result from commercial software.
Results: A series of the light/radiation field coincidence test and the Winston-Lutz test were analyzed using the developed device and software. The results could be easily confirmed using the reporting function of the software. Regarding the result of the Winston-Lutz test, most of the analysis differences between the developed software and commercially available software were within the pixel size (0.22 mm).
Discussions: Since the accuracy of the radiation field affects the result of the Winston-Lutz test, the presented procedure of performing the light/radiation coincidence field test in advance facilitates the interpretation of the error of the Winston-Lutz test. Based on the results of the Winston-Lutz test, we were able to demonstrate the tuning flow of all in-room lasers including the horizontality and verticality by using the developed device.
Conclusions: We have developed a couch-mounted verification device and software that can evaluate the light/radiation coincidence field test and the alignment including the horizontality and verticality of the in-room laser used for patient positioning in radiotherapy, and reported its usefulness. The analysis accuracy of the developed software was comparable to that of commercially available software. The use of this device and the developed software would contribute to not only the efficiency of adjusting all in-room lasers' alignment including the horizontality and verticality but also reflect accurately the result of the Winston-Lutz test.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.11323/jjmp.41.1_1 | DOI Listing |
Front Oncol
October 2024
Department of Radiation Oncology, School of Medicine, Stanford University, Stanford, CA, United States.
J Med Phys
June 2024
Department of Radiation Oncology, Sir H.N Reliance Foundation and Research Centre, Mumbai, Maharashtra, India.
Recent decades have witnessed transformative advances in radiation physics and computer technology, revolutionizing the precision of radiation therapy. The adoption of intricate treatment techniques such as three-dimensional conformal radiotherapy, intensity-modulated radiotherapy, volumetric-modulated arc therapy, and image-guided radiotherapy necessitates robust quality assurance (QA) programs. This study introduces the SunCHECK™ Machine (SCM), a web-based QA platform, presenting early results from its integration into a comprehensive QA program.
View Article and Find Full Text PDFJ Appl Clin Med Phys
October 2024
Radiation oncology department, Texas Oncology, Houston, Texas, USA.
Purpose: The aim of this study is to find optimal gantry, collimator, and couch angles for performing single isocenter, multiple target stereotactic radiosurgery (SIMT-SRS). Nineteen angle sets were tested across seven linear accelerators for radiation-isocenter coincidence and off-isocenter coincidence. The off-isocenter Winston-Lutz test was performed to evaluate the accuracy of isocenter alignment for each angle set, and optimal angle sets as well as maximum off-isocenter distance to target for each angle set was determined.
View Article and Find Full Text PDFPhys Med
April 2024
Radiation Therapy Department, CHR Metz-Thionville, Metz, France.
Purpose: Stereotactic radiotherapy (SRT) has transformed cancer treatment, especially for brain metastases. Ensuring accurate SRT delivery is crucial, with the Winston-Lutz test being an important quality control tool. Off-axis Winston-Lutz (OAWL) tests are designed for accuracy assessment, but most are limited to fixed angles and hampered by local-field shifts caused by suboptimal Multi-Leaf Collimator (MLC) positioning.
View Article and Find Full Text PDFDiagnostics (Basel)
February 2024
Department of Radiation Oncology, Yeungnam University Medical Center, Daegu 42415, Republic of Korea.
HyperArc is a preferred technique for treating brain metastases, employing a single isocenter for multiple lesions. Geometrical isocentricity in the TrueBeam linear accelerator with HyperArc is crucial. We evaluated machine performance checks (MPCs) as an alternative to the Winston-Lutz (WL) test to verify the treatment isocenter.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!