Purpose: To investigate the possibility of reducing the injected activity for whole-body [18F]FDG-PET/CT studies of paediatric oncology patients and to assess the usefulness of time-of-flight (TOF) acquisition on PET image quality at reduced count levels.

Procedures: Twenty-nine paediatric oncology patients (12F/17M, 3-18 years old (median age 13y), weight 45±20 kg, BMI 19±4 kg/m), who underwent routine whole-body PET/CT examinations on a Siemens Biograph mCT TrueV system with TOF capability (555ps) were included in this study. The mean injected activity was 156 ± 45 MBq (3.8 ± 0.8 kg/MBq) and scaled to patient weight. The raw data was collected in listmode (LM) format and pre-processed to simulate reduced levels of [18F]FDG activity (75, 50, 35, 20 and 10% of the original counts) by randomly removing events from the original LM data. All data were reconstructed using the vendor-specific e7-tools with standard OSEM only, with OSEM plus resolution recovery (PSF). The reconstructions were repeated with added TOF (TOF) and PSF+TOF. The benefit of TOF together with the reduced count levels was evaluated by calculating the gains in signal-to-noise ratio (SNR) in the liver and contrast-to-noise ratio (CNR) in all PET-positive lesions before and after TOF employed at every simulated reduced count level. Finally, the PSF+TOF images at 50, 75 and 100% of counts were evaluated clinically on a 5-point scale by three nuclear medicine physicians.

Results: The visual inspection of the reconstructed images did not reveal significant differences in image quality between 75 and 100% count levels for PSF+TOF. The improvements in SNR and CNR were the greatest for TOF reconstruction and PSF combined. Both SNR and CNR gains did increase linearly with the patients BMI for both OSEM only and PSF reconstruction. These benefits were observed until reducing the counts to 50 and 35% for SNR and CNR, respectively.

Conclusions: The benefit of using TOF was noticeable when using 50% or greater of the counts when evaluating the CNR and SNR. For [18F]FDG-PET/CT, whole-body paediatric imaging the injected activity can be reduced to 75% of the original dose without compromising PET image quality.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC8410733PMC
http://dx.doi.org/10.1007/s11307-021-01601-4DOI Listing

Publication Analysis

Top Keywords

injected activity
12
image quality
12
reduced count
12
snr cnr
12
paediatric oncology
8
oncology patients
8
tof
8
pet image
8
benefit tof
8
count levels
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!