Objectives: Opening-wedge high tibia osteotomies (HTO) can be technically challenging. The HTO iBalance system was designed to reduce vascular complications and to avoid secondary plate removal. The purpose of the study was to evaluate the performance of the HTO iBalance system in patients with symptomatic medial osteoarthritis and varus malalignment.

Methods: The study was performed as a retrospective cohort study investigating a consecutive series of patients who underwent HTO with the iBalance system performed by a single surgeon from August 2013 to March 2016 at Zealand University Hospital, Koege, and Aleris-Hamlet Hospital. The primary outcome was the degree of realignment. The secondary outcome was Knee injury and Osteoarthritis Outcome Score (KOOS). Follow-up was performed at mean (SD) 25 (9.7) months. Weight-bearing long-leg standing radiographs were taken before surgery and at follow-up. Failure was defined as collapse of the HTO defined as a correction <50% of the intended correction at time of follow-up. Logistic regression was used to identify risk factors for failure.

Results: 44 patients and a total of 47 knees were included in this study. Preoperatively the mechanical axis was a mean (SD) 5.8° (2.9) varus and postoperatively 2.3° (3.7) varus . The HTO failed in 13 of 47 knees (28%). Patients with failure showed no statistically significant differences to non-failure in any KOOS subscore (p>0.05). American Society of Anesthesiologists (ASA) score (p=0.01) and body mass index (BMI) (p=0.05) were correlated with failure, whereas bone transplantation and smoking were not.

Conclusion: In this study, the failure rate of HTO was 28%. High BMI and ASA-score were the only risk factors associated with failure while bone grafting and smoking were not.

Level Of Evidence: Retrospective cohort study, level III.

Download full-text PDF

Source
http://dx.doi.org/10.1136/jisakos-2019-000394DOI Listing

Publication Analysis

Top Keywords

hto ibalance
12
ibalance system
12
failure rate
8
retrospective cohort
8
cohort study
8
failure bone
8
hto
6
failure
5
study
5
high tibial
4

Similar Publications

Objectives: Opening-wedge high tibia osteotomies (HTO) can be technically challenging. The HTO iBalance system was designed to reduce vascular complications and to avoid secondary plate removal. The purpose of the study was to evaluate the performance of the HTO iBalance system in patients with symptomatic medial osteoarthritis and varus malalignment.

View Article and Find Full Text PDF

Purpose: High tibial osteotomy (HTO) is an invaluable tool in the treatment of a variety of conditions in active patients with varus malalignment of the knee. The polyetheretherketone (PEEK) iBalance HTO system has been designed to make the osteotomy safer, more reproducible, and avoid metal hardware related complications. The purpose of this study was to evaluate short-term outcomes in patients who underwent an open wedge high tibial osteotomy with the iBalance HTO system.

View Article and Find Full Text PDF

Background: This study aimed to assess the mechanical static and fatigue strength provided by the FlexitSystem plate in medial opening wedge high tibial osteotomies (MOWHTO), and to compare it to six previously tested implants: the TomoFix small stature, the TomoFix standard, the ContourLock, the iBalance, the second generation PEEKPower and the size 2 Activmotion. Thus, this will provide surgeons with data that will help in the choice of the most appropriate implant for MOWHTO.

Methods: Six fourth-generation tibial bone composites underwent a MOWHTO and each was fixed using six FlexitSystem plates, according to standard techniques.

View Article and Find Full Text PDF

Background: Many different fixation devices are used to maintain the correction angle after medial open wedge high tibial osteotomy (MOWHTO). Each device must provide at least sufficient mechanical stability to avoid loss of correction and unwanted fracture of the contralateral cortex until the bone heals. In the present study, the mechanical stability of following different implants was compared: the TomoFix small stature (sm), the TomoFix standard (std), the Contour Lock, the iBalance and the second generation PEEKPower.

View Article and Find Full Text PDF

Background: High tibial osteotomy (HTO) is a valuable treatment option in the high-demand patient with chondral damage and an altered mechanical axis. Traditional opening wedge HTO performed with metal plates has several limitations, including hardware irritation, obscuration of detail on magnetic resonance imaging, and complexity of revision surgery. Recently, an all-polyetheretherketone (PEEK) HTO implant was introduced, but no studies to date have evaluated the performance of this implant with minimum 2-year outcomes compared with a traditional metal plate.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!