Background: Although the Pinnacle Acetabular Hip System (DePuy Synthes) has demonstrated excellent survivorship results since it was first introduced in 2003, there have been a growing number of cases indicating that Pinnacle liners may be subject to a higher-than-expected rate of early dissociation failure. Between 2006 and 2020, our Centre received 212 retrieved Pinnacle liners from Western Australian hospitals. Of these, 26 were removed due to liner dissociation.
Methods: To better understand the frequency and cause of this complication we assessed all retrieved Pinnacle acetabular components for type, damage modes and patient demographics. The leverage force required to dissociate Pinnacle liners was also measured and compared with another commonly used acetabular system, the Trident (Stryker Orthopaedics).
Results: The estimated minimum incidence of liner dissociation from our data was 0.35%. Characterisation of dissociated Pinnacle cases (n = 26) revealed 73% were female with an average age of 59 compared to all retrieved Pinnacle cases (n = 212) where 58% were female with an average age of 66. Retrieval analysis indicated plastic deformation of the liner into an ovoid shape, signs of impingement on the rim postero-superiorly and shearing of the liner's anti-rotation tabs was common. Mechanical testing indicated that the dissociation strength of Pinnacle cups decreases at approximately 6.6 N/year in situ (p = 0.01).
Conclusion: The survival rate of Pinnacle acetabular cups is exceptional with only 5% revised at 10 years. However, surgeons should be aware of the clinical symptoms and high-risk demographics when assessing patients with polyethylene Pinnacle liners.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1177/11207000211008459 | DOI Listing |
Technol Health Care
September 2024
Hospital of Trauma Surgery, Marktredwitz Hospital, Marktredwitz, Germany.
Background: Uncemented acetabular components are widely used in modern total hip arthroplasty (THA). Modularity has numerous advantages including the ability to use supplementary screw fixation for the shell, and allow to switch from ceramic to polyethylene (PE) bearings and vice versa, and the use of lipped and face-changing liners.
Objective: Despite these advantages, a problem with modular PE liners is dissociation.
J Arthroplasty
July 2023
Department of Orthopedic Surgery, Mayo Clinic, Rochester, Minnesota.
Background: The advent of highly porous ingrowth surfaces and highly crosslinked polyethylene has been expected to improve implant survivorship in revision total hip arthroplasty. Therefore, we sought to evaluate the survival of several contemporary acetabular designs following revision total hip arthroplasty.
Methods: Acetabular revisions performed from 2000 to 2019 were identified from our institutional total joint registry.
J Orthop Res
August 2023
Clemson University - MUSC Bioengineering Program, Charleston, South Carolina, USA.
Acetabular components (DePuy Pinnacle (A) and Stryker Trident (B), Ti-6Al-4V shells and CoCrMo liners) with varying geometries were assembled under a 4 kN seating load. Liner-displacement was recorded. Cyclic compression to 4 kN, R = 0.
View Article and Find Full Text PDFBMC Musculoskelet Disord
August 2022
Department of Orthopedic Surgery, Nanfang Hospital, Southern Medical University, Guangzhou, China.
Background: Total hip arthroplasty(THA)is widely used to treat end-stage hip disorders. Ceramic-on-ceramic total hip prostheses are widely used because of their durability. Alumina matrix composite (AMC), known as the fourth-generation ceramics, reduces implant fracture and wear rate compared to their predecessors.
View Article and Find Full Text PDFArch Orthop Trauma Surg
May 2023
Laboratory of Biomechanics and Implant Research, Center for Orthopedics, Trauma Surgery and Spinal Cord Injury, Heidelberg University Hospital, Heidelberg, Germany.
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!