Dissociation of polyethylene liners with the Depuy Pinnacle cup: a report of 26 cases.

Hip Int

Department of Medical Engineering and Physics, Centre for Implant Technology and Retrieval Analysis, Royal Perth Hospital, Perth, Australia.

Published: January 2023

AI Article Synopsis

  • The Pinnacle Acetabular Hip System has shown strong long-term success since 2003, but recent cases indicate a concerning rate of early dissociation failures, especially in women.
  • Between 2006 and 2020, 212 Pinnacle liners were retrieved from patients, with 26 cases specifically linked to liner dissociation, revealing characteristics like significant plastic deformation in the liners.
  • The study found a minimum incidence of 0.35% for liner dissociation, suggesting that while the overall survival rate is high, awareness of symptoms and at-risk patient demographics is crucial for surgeons.

Article Abstract

Background: Although the Pinnacle Acetabular Hip System (DePuy Synthes) has demonstrated excellent survivorship results since it was first introduced in 2003, there have been a growing number of cases indicating that Pinnacle liners may be subject to a higher-than-expected rate of early dissociation failure. Between 2006 and 2020, our Centre received 212 retrieved Pinnacle liners from Western Australian hospitals. Of these, 26 were removed due to liner dissociation.

Methods: To better understand the frequency and cause of this complication we assessed all retrieved Pinnacle acetabular components for type, damage modes and patient demographics. The leverage force required to dissociate Pinnacle liners was also measured and compared with another commonly used acetabular system, the Trident (Stryker Orthopaedics).

Results: The estimated minimum incidence of liner dissociation from our data was 0.35%. Characterisation of dissociated Pinnacle cases (n = 26) revealed 73% were female with an average age of 59 compared to all retrieved Pinnacle cases (n = 212) where 58% were female with an average age of 66. Retrieval analysis indicated plastic deformation of the liner into an ovoid shape, signs of impingement on the rim postero-superiorly and shearing of the liner's anti-rotation tabs was common. Mechanical testing indicated that the dissociation strength of Pinnacle cups decreases at approximately 6.6 N/year in situ (p = 0.01).

Conclusion: The survival rate of Pinnacle acetabular cups is exceptional with only 5% revised at 10 years. However, surgeons should be aware of the clinical symptoms and high-risk demographics when assessing patients with polyethylene Pinnacle liners.

Download full-text PDF

Source
http://dx.doi.org/10.1177/11207000211008459DOI Listing

Publication Analysis

Top Keywords

pinnacle liners
16
pinnacle acetabular
12
retrieved pinnacle
12
pinnacle
11
pinnacle cases
8
female average
8
average age
8
liners
5
dissociation
4
dissociation polyethylene
4

Similar Publications

Background: Uncemented acetabular components are widely used in modern total hip arthroplasty (THA). Modularity has numerous advantages including the ability to use supplementary screw fixation for the shell, and allow to switch from ceramic to polyethylene (PE) bearings and vice versa, and the use of lipped and face-changing liners.

Objective: Despite these advantages, a problem with modular PE liners is dissociation.

View Article and Find Full Text PDF

Background: The advent of highly porous ingrowth surfaces and highly crosslinked polyethylene has been expected to improve implant survivorship in revision total hip arthroplasty. Therefore, we sought to evaluate the survival of several contemporary acetabular designs following revision total hip arthroplasty.

Methods: Acetabular revisions performed from 2000 to 2019 were identified from our institutional total joint registry.

View Article and Find Full Text PDF

Acetabular components (DePuy Pinnacle (A) and Stryker Trident (B), Ti-6Al-4V shells and CoCrMo liners) with varying geometries were assembled under a 4 kN seating load. Liner-displacement was recorded. Cyclic compression to 4 kN, R = 0.

View Article and Find Full Text PDF

Background: Total hip arthroplasty(THA)is widely used to treat end-stage hip disorders. Ceramic-on-ceramic total hip prostheses are widely used because of their durability. Alumina matrix composite (AMC), known as the fourth-generation ceramics, reduces implant fracture and wear rate compared to their predecessors.

View Article and Find Full Text PDF

Dissociation of liner from cup in THA: does liner damage affect the risk of dissociation?

Arch Orthop Trauma Surg

May 2023

Laboratory of Biomechanics and Implant Research, Center for Orthopedics, Trauma Surgery and Spinal Cord Injury, Heidelberg University Hospital, Heidelberg, Germany.

Article Synopsis
  • The study looked at a problem with hip replacement parts coming apart, especially in a system called Pinnacle.
  • Researchers tested how strong different hip replacement parts were and which ones stayed together best under pressure.
  • They found that Pinnacle parts were the strongest when they had a safety feature, but if that feature was removed, they were the easiest to come apart.
View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!