A study of dose indicators during intra-arterial catheter implantation for liver chemotherapy.

J Radiol Prot

Department of Radiology, Institut Paoli Calmettes, 232 Boulevard Sainte Marguerite, Marseille 13009, France.

Published: August 2021

AI Article Synopsis

Article Abstract

The aim of the present study was to describe patient dose indicator levels during intra-arterial catheter (IAC) implantation for liver chemotherapy, and to determine factors affecting the dose indicators. Between January 2017 and January 2019, 61 patients with hepatic metastases from colorectal cancer were retrospectively included. Interventions were carried out in a standardised manner by three experienced radiologists on the same angiographic table without changes in protocol parameters. For each patient, clinical, radiological and dosimetry data were collected, including the air kerma area product (KAP), part of KAP due to the fluoroscopy and fluoroscopy time (FT), total kerma at the reference interventional point and peak skin dose (PSD). Local dose reference levels (RLs) were determined as the third quartile of the patient dose distributions. Univariate and multivariate analysis of factors affecting dose indicators was performed. The mean KAP was 111 Gy cm, the mean reference point air kerma (Ka,r) was 648 mGy, the mean PSD was 613 mGy, and the mean FT was 3190 s (62% of the KAP). The mean cone beam computed tomography dose was 37.3 ± 11.8 Gy cm, which accounted for 37% of the KAP. The RL could be proposed taking into account the third quartiles (KAP = 164.6 Gy cm, Ka,r = 904.5 mGy, FT = 4011 s and standard deviation = 772.7 mGy). The factors affecting dose indicators were related to the patients (sex, cardiovascular risk factors, weight, body mass index), to the vascular anatomies (coeliac trunk angulation) and to the procedures (number of embolised arteries). This study allowed a better understanding of dose indicators and factors affecting these indicators during the implantation of IACs for hepatic chemotherapy, which is a long and difficult procedure. Local dose RLs were determined. Multicentre, multi-equipment studies are necessary.

Download full-text PDF

Source
http://dx.doi.org/10.1088/1361-6498/abf570DOI Listing

Publication Analysis

Top Keywords

dose indicators
20
factors dose
12
dose
10
intra-arterial catheter
8
implantation liver
8
liver chemotherapy
8
patient dose
8
air kerma
8
local dose
8
rls determined
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!