Objectives: We analyzed the effects of the Er:YAG laser used with different parameters on dentinal tubule (DT) occlusion, intrapulpal temperature and pulp tissue morphology in order to determine the optimal parameters for treating dentin hypersensitivity.
Methodology: Dentin specimens prepared from 36 extracted human third molars were randomized into six groups according to the treatment method (n=6 each): control (A); Gluma desensitizer (B); and Er:YAG laser treatment at 0.5 W , 167 J/cm2 (50 mJ, 10 Hz) (C), 1 W , 334 J/cm2 (50 mJ, 20 Hz) (D), 2 W , 668 J/cm2 (100 mJ, 20 Hz) (E), and 4 W and 1336 J/cm2 (200 mJ, 20 Hz) (F). Treatment-induced morphological changes of the dentin surfaces were assessed using scanning electron microscopy (SEM) to find parameters showing optimal dentin tubule occluding efficacy. To further verify the safety of these parameters (0.5 W, 167 J/cm2), intrapulpal temperature changes were recorded during laser irradiation, and morphological alterations of the dental pulp tissue were observed with an upright microscope.
Results: Er:YAG laser irradiation at 0.5 W (167 J/cm2) were found to be superior in DT occlusion, with an exposure rate significantly lower than those in the other groups (P<0.05). Intrapulpal temperature changes induced by Er:YAG laser irradiation at 0.5 W (167 J/cm2) with (G) and without (H) water and air cooling were demonstrated to be below the threshold. Also, no significant morphological alterations of the pulp and odontoblasts were observed after irradiation.
Conclusion: Therefore, 0.5 W (167 J/cm2) is a suitable parameter for Er:YAG laser to occlude DTs, and it is safe to the pulp tissue.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9648964 | PMC |
http://dx.doi.org/10.1590/1678-7757-2020-0266 | DOI Listing |
J Prosthodont Res
January 2025
Department of Orthodontics, Osaka Dental University, Hirakata, Japan.
Purpose: To perform vertical bone augmentation on rat parietal bone by coating the inner surface of dense polytetrafluoroethylene (d-PTFE) domes with hydroxyapatite (HA) using Erbium Yttrium Aluminum Garnet (Er:YAG) pulsed laser deposition in a rat model.
Methods: The d-PTFE plate surface, α-tricalcium phosphate (α-TCP) coating, and HA coating were measured using scanning electron microscopy and X-ray diffraction to confirm the replacement of α-TCP with HA via high-pressure steam sterilization. The dome was glued to the center of the rat parietal bone and closed with periosteal and epithelial sutures.
Heliyon
November 2024
Department of Orthodontics, Faculty of Dentistry, University of Health Sciences, Turkey.
Aim: To evaluate the shear bond strength (SBS) of ceramic and metallic orthodontic brackets bonded to lithium disilicate ceramics or hybrid ceramics and subjected to different surface conditioning treatments.
Materials And Methods: In total, 300 specimens were fabricated from GC LiSi (lithium disilicate) and GC Cerasmart (hybrid) ceramic blocks. The specimens were divided into four groups according to the following surface treatments: hydrofluoric acid (HF); sandblasting with 50 μm aluminum oxide; Monobond Etch and Prime; and erbium-doped yttrium aluminum garnet (Er-YAG) laser.
Braz Dent J
July 2024
Conservative Dentistry Department, Faculty of Dentistry, Cairo University, Egypt. Postal Code: 11553.
ER:YAG laser and experimental resin-based dental adhesive loaded with functionalized carbonated apatite filler were used in this study to evaluate the dentin interaction in terms of penetration and occlusion of the dentinal tubules aiding in the control of dentin hypersensitivity (DH). Spheroidal Carbonated apatite nanoparticles (N-CAP), with an average size of 20±5 nm diameter, were synthesized, characterized, and incorporated in a universal adhesive "All Bond Universal, Bisco, USA", in (2% weight) concentration. Er:YAG laser "Lightwalker, FOTONA, EU" was adjusted to an energy output of 40mJ/ pulse and pulse repetition of 10 Hz for 10 seconds.
View Article and Find Full Text PDFLasers Med Sci
April 2024
Department of Periodontology, Hamidiye Dental Faculty, University of Health Sciences, Istanbul, Turkey.
Aim: The aim of the present study was to evaluate the efficacy of 30-angled Er:YAG laser tip and different periodontal instruments on root surface roughness and morphology in vitro.
Methods: Eighteen bovine teeth root without carious lesion were decoronated from the cementoenamel junction and seperated longitidunally. A total of 36 obtained blocks were mounted in resin blocks and polished with silicon carbide papers under water irrigation.
J Lasers Med Sci
November 2023
School of Dentistry, University of São Paulo, São Paulo, São Paulo, Brazil.
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!