Background: The quantification of mitochondrial DNA heteroplasmy for the diagnosis of mitochondrial disease or after mitochondrial donation, is performed mainly using next-generation sequencing strategies (NGS). Digital PCR (dPCR) has the potential to offer an accurate alternative for mutation load quantification.
Methods: We assessed the mutation load of 23 low-input human samples at the m.11778 locus, which is associated with Leber's hereditary optic neuropathy (LHON) using 2 droplet digital PCR platforms (Stilla Naica and Bio-Rad QX200) and the standard NGS strategy. Assay validation was performed by analyzing a titration series with mutation loads ranging from 50% to 0.01%.
Results: A good concordance in mutation rates was observed between both dPCR techniques and NGS. dPCR established a distinctly lower level of background noise compared to NGS. Minor alleles with mutation loads lower than 1% could still be detected, with standard deviations of the technical replicates varying between 0.07% and 0.44% mutation load. Although no significant systematic bias was observed when comparing dPCR and NGS, a minor proportional bias was detected. A slight overestimation of the minor allele was observed for the NGS data, most probably due to amplification and sequencing errors in the NGS workflow.
Conclusion: dPCR has proven to be an accurate tool for the quantification of mitochondrial heteroplasmy, even for samples harboring a low mutation load (<1%). In addition, this alternative technique holds multiple benefits compared to NGS (e.g., less hands-on time, more straightforward data-analysis, and a lower up-front capital investment).
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1093/clinchem/hvab021 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!