Objective: To evaluate a negative pressure microenvironment designed to contain laser plume during flexible transnasal laryngoscopy.

Methods: The Negative Pressure Face Shield (NPFS) was previously reported as well tolerated with initial use on 30 patients. Diagnostic transnasal laryngoscopy was performed on an additional 108 consecutive patients who were evaluated by questionnaires and sequential pulse oximetry. Further study addressed operative transnasal potassium-titanyl-phosphate (KTP) laser laryngoscopy with biopsy done on four patients employing the NPFS.

Results: The previously described NPFS version 3 (v.3), a transparent acrylic barrier with two anterior instrumentation ports, was modified by repositioning the side suction port closer to the level of the nose and deepening the lateral sides, squaring off the lower projection. A post-procedure questionnaire employing a 5-point Likert scale ranging from no symptoms (rating of 1) to intolerable (rating of 5) identified excellent patient tolerance of the new design (v.4), among 22 patients evaluated and similar in the comparison to the 116 patients using version 3. Among the 138 patients analyzed, only one patient rated the experience as greater than "mild claustrophobia." 100% of patients answered either "none" or "mild" to the pain and shortness of breath questions. The NPFS (v.4) was then successfully used in four patients for laser laryngoscopy with biopsy of laryngeal papilloma (3/4) and hemorrhagic polyp (1/4). Post-procedure questionnaire identified no shortness of breath (4/4), no claustrophobia (4/4), no pain (4/4) and no significant changes in pulse oximetry during use.

Conclusion: Extensive experience in performing diagnostic laryngoscopy with the NPFS directed design changes leading to successful use for transnasal flexible laser laryngoscopy with biopsy in a negative pressure microenvironment.

Level Of Evidence: Level 2b (Cohort Study).

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC8013799PMC
http://dx.doi.org/10.1002/lio2.526DOI Listing

Publication Analysis

Top Keywords

negative pressure
12
laser laryngoscopy
12
laryngoscopy biopsy
12
laser plume
8
flexible transnasal
8
transnasal laryngoscopy
8
patients
8
patients evaluated
8
pulse oximetry
8
post-procedure questionnaire
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!