Absolute calibration of the Lyman-α measurement apparatus at DIII-D.

Rev Sci Instrum

Princeton Plasma Physics Laboratory, Princeton, New Jersey 08543, USAMIT Plasma Science and Fusion Center, Cambridge, Massachusetts 02139, USAOak Ridge National Laboratory, Oak Ridge, Tennessee 37831, USA.

Published: March 2021

The LLAMA (Lyman-Alpha Measurement Apparatus) diagnostic was recently installed on the DIII-D tokamak [Rosenthal et al., Rev. Sci. Instrum. (submitted) (2020)]. LLAMA is a pinhole camera system with a narrow band Bragg mirror, a bandpass interference filter, and an absolute extreme ultraviolet photodiode detector array, which measures the Ly-α brightness in the toroidal direction on the inboard, high field side (HFS) and outboard, low field side (LFS). This contribution presents a setup and a procedure for an absolute calibration near the Ly-α line at 121.6 nm. The LLAMA in-vacuum components are designed as a compact, transferable setup that can be mounted in an ex situ vacuum enclosure that is equipped with an absolutely calibrated Ly-α source. The spectral purity and stability of the Ly-α source are characterized using a vacuum ultraviolet spectrometer, while the Ly-α source brightness is measured by a NIST-calibrated photodiode. The non-uniform nature of the Ly-α source emission was overcome by performing a calibration procedure that scans the Ly-α source position and employs a numerical optimization to determine the emission pattern. Nominal and measured calibration factors are determined and compared, showing agreement within their uncertainties. A first conversion of the measured signal obtained from DIII-D indicates that the Ly-α brightness on the HFS and LFS is on the order of 10 Ph sr m s. The established calibration setup and procedure will be regularly used to re-calibrate the LLAMA during DIII-D vents to monitor possible degradation of optical components and detectors.

Download full-text PDF

Source
http://dx.doi.org/10.1063/5.0038134DOI Listing

Publication Analysis

Top Keywords

ly-α source
20
absolute calibration
8
measurement apparatus
8
ly-α
8
ly-α brightness
8
field side
8
setup procedure
8
source
5
calibration lyman-α
4
lyman-α measurement
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!