A new technique to attenuate the unshifted deuterium Balmer-alpha (D-alpha) emission is developed and tested for the fast ion D-alpha (FIDA) diagnostic. The unshifted D-alpha emission, at λ = 656.1 nm, is around three orders of magnitude higher than the desired FIDA emission. Blocking the strong emission feature is essential to prevent blooming and light smearing on the CCD chip and scattered light contamination. The new method is a notch filter approach that utilizes the reflection from ultra-narrow bandpass filters to block the saturating signal before it enters the spectrometer. Collimated light from the fibers is reflected off the filter at a 15° angle of incidence. Measurements show that a center wavelength transmission of 0.006 and a blocking full width at half maximum of ∼1 nm are achieved by using a 200 μm fiber and a 20 mm focal length collimator with two filters.