AI Article Synopsis

  • This study examines radiation exposure in pediatric patients undergoing transcatheter closure for ventricular septal defects (VSD) and analyzes various factors impacting that exposure.* -
  • Over a span of three years, it reviewed 85 cases, revealing that higher patient weight, more operators involved, complex device positioning, and procedural incidents significantly increased radiation exposure.* -
  • The findings highlight the need for careful management of these factors to minimize radiation doses in children during interventional cardiology procedures.*

Article Abstract

Background: Ventricular septal defect (VSD) transcatheter closure is gaining worldwide popularity despite its complexity. Reports on key factors affecting radiation exposure in children are scarce.

Aims: This clinical study is the first to comprehensively analyze the impact of all relevant parameters on children's radiation exposure during VSD interventional closures.

Methods: Between March 2016 and August 2019, all pediatric VSD cases percutaneously treated at a reference center for interventional congenital cardiology and equipped with a single-plane Innova 2100 X-ray unit were retrospectively reviewed. Multiple linear regression was performed to investigate the impact of clinical, technical, and procedural parameters on patients' radiation exposure assessed using total air kerma area product (P).

Results: A total of 85 patients were included in this study and 82.4% had perimembranous defects. Device implantation was successful in 96.5% of cases. The procedure lasted for a median of 60 min with a median P of 19.6 Gy.cm (range, 1.1 to 244.8 Gy.cm). Patients' weight (B = 1.679, P = 0.01), number of operators (B = 1.561, P = 0.02), device positioning complexity (B = 2.381, P = 0.002), and procedural incidents (B = 2.096, P = 0.008) significantly increased P. Patients' age (B = 1.053, P = 0.784), device design (B = -1.216, P = 0.780) and approach of delivery (B = -1.119, P = 0.511) did not significantly affect P.

Conclusions: Radiation exposure in children undergoing VSD percutaneous closure was highly variable. A higher patient's weight, numbers of operators, complexity in device positioning, and procedural incidents, were identified as key factors increasing patient dose for this kind of intervention.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC8012278PMC

Publication Analysis

Top Keywords

radiation exposure
20
percutaneous closure
8
ventricular septal
8
key factors
8
exposure children
8
device positioning
8
procedural incidents
8
radiation
5
exposure
5
closure ventricular
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!