Background: The ADVIA 120 is a widely used hematology analyzer, which has not been previously validated for determining differential leukocyte counts in sheep.
Objectives: We aimed to compare differential leukocyte counts on the ADVIA 120 (A-Diff) with counts obtained using the manual method (M-Diff) in sheep.
Methods: Ethylenediaminetetraacetic acid-anticoagulated blood samples analyzed within 4 hours of collection were used. Samples with inappropriately filled tubes, detectable clots, overtly erroneous ADVIA peroxidase cytograms, and poor-quality blood smears were excluded from the study. Two independent observers compared the results of the A-Diff with those of the M-Diff. The M-diff was performed by counting 200 leukocytes on a blood smear.
Results: Overall, 88 samples (44 rams and 44 ewes) were included. The correlation between the A-Diff and M-Diff was high for neutrophils (r = .873, P < .001), lymphocytes (r = .863, P < .001), and eosinophils (r = .750, P < .001), and low for monocytes (r = .212, P = .048). The Passing-Bablok regression analyses revealed constant error for eosinophils [1.17%; 95% confidence interval (CI): 0.67%, 1.55%] and proportional error for lymphocytes (0.84; 95% CI: 0.74, 0.95) and eosinophils (0.85; 95% CI: 0.74, 0.96). The Bland-Altman analyses revealed negative biases of 2.4% and 3.0% for neutrophils and lymphocytes, respectively, and positive biases of 3.2% and 0.8% for monocytes and eosinophils, respectively. The extremely low basophil percentages precluded a meaningful method comparison for this leukocyte type.
Conclusions: The ADVIA 120 appears to perform well for determining neutrophil, lymphocyte, and eosinophil percentages in sheep as compared with the manual method. However, blood smear examinations are recommended for the confirmation of abnormal differential leukocyte count results.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1111/vcp.12935 | DOI Listing |
Background: The aetiology of increased mean platelet volume in dogs is currently uncertain. Our aim was to investigate the aetiology of increased mean platelet volume in dogs with and without thrombocytopenia.
Methods: The database of a teaching hospital was retrospectively searched for dogs with increased mean platelet volume (>14.
Vet Clin Pathol
March 2024
Diagnostic Laboratory, School of Veterinary Medicine, Faculty of Health Sciences, Aristotle University of Thessaloniki, Thessaloniki, Greece.
Background: Although widely used, the ADVIA 120 hematology analyzer has not been previously validated for determining the differential leukocyte count in goats.
Objectives: The aim of this study was to compare the differential leukocyte counts provided by the ADVIA 120 (A-diff) and the manual method (M-Diff) in goats.
Methods: EDTA blood samples that were analyzed within 4 h of collection were used in the study.
Vet Clin Pathol
March 2024
Pathology Department, Lilly Research Laboratories, Eli Lilly and Company, Lilly Corporate Center, Indianapolis, Indiana, USA.
Background: The Sysmex XN-1000V automated hematology analyzer with multispecies software was released in June 2017 for use in research laboratories. Laser light, impedance, fluorescent staining, and fluorescent flow cytometry are used to analyze whole blood for CBC, reticulocyte counts, and WBC counts, including a 5-part differential leukocyte analysis.
Objectives: A side-by-side comparison of the Sysmex XN-1000V with the Siemens ADVIA 120 in analyzing blood from healthy mice and rats will provide insight into the performance of the new analyzer and its capabilities for use in drug development studies.
J Avian Med Surg
November 2022
Department of Pathology, Microbiology, and Immunology, University of California Davis, School of Veterinary Medicine, Davis, CA 95616, USA.
Although cockatiels are among the most common avian species maintained as companion animals in the United States, information on standard hematologic reference values for this species is limited. The objectives of this study were to establish hematologic reference intervals (RI) for cockatiels, compare methods using both the Natt-Herrick technique (NHT) and the smear-based estimation technique (SBT), explore age and sex differences in the hematologic findings for this species, and produce the first cockatiel RI for fibrinogen concentration and thrombocyte estimate. Healthy cockatiels (60 males and 60 females, 2-11 years old) from a research colony were included in this study.
View Article and Find Full Text PDFVet Clin Pathol
March 2023
Clinic for Small Animals, Internal Medicine, Department of Veterinary Clinical Sciences, Faculty of Veterinary Medicine, Justus Liebig University, Giessen, Germany.
Background: A novel method using a moving threshold (r-PLTmt) to determine canine r-PLTs (reticulated platelets) has been introduced for ADVIA 2120 software v6.11.7.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!