There is a need for accurate diagnostic tests for severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2), the cause of coronavirus disease (COVID-19). This study aimed to evaluate the diagnostic accuracy of an immunochromatography-based immunoglobulin G (IgG)/immunoglobulin M (IgM) antibody assay (GenBody™ COVI040) for detecting SARS-CoV-2 antibody seroconversion in COVID-19 patients. A total of 130 samples, serially collected from patients with confirmed COVID-19, and 100 negative control samples were tested for anti-SARS-CoV-2 IgM and IgG using the GenBody™ COVI040 assay following the South Korean Ministry of Food and Drug Safety guidelines on the review and approval of in vitro diagnostic devices for COVID-19. Reverse-transcription polymerase chain reaction results were used as the comparator. The overall sensitivity of the GenBody™ COVI040 assay was 97.69% (95% confidence interval (CI): 93.40-99.52%). The sensitivity of the assay increased with time post symptom onset (PSO) (sensitivity ≤6 days PSO: 78.57%, 95% CI: 49.20-95.34%; sensitivity 7-13 days PSO: 100%, 95% CI: 87.23-100%; and sensitivity ≥14 days PSO: 100%, 95% CI: 95.94-100%). The specificity of the assay was 100% (95% CI: 96.38-100%). The GenBody™ COVI040 assay showed high sensitivity and specificity, making it a promising diagnostic test to monitor COVID-19.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC8002734PMC
http://dx.doi.org/10.3390/diagnostics11030537DOI Listing

Publication Analysis

Top Keywords

genbody™ covi040
20
covi040 assay
12
days pso
12
100% 95%
12
antibody assay
8
assay genbody™
8
antibody seroconversion
8
pso 100%
8
assay
7
sensitivity
6

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!