Background: High-sensitivity cardiac troponin (hs-cTn) assays have different analytic characteristics.
Objectives: The goal of this study was to quantify differences between assays for common analytical benchmarks and to determine whether they may result in differences in the management of patients with suspected acute coronary syndrome (ACS).
Methods: The authors included patients with suspected ACS enrolled in the ROMICAT (Rule Out Myocardial Infarction/Ischemia Using Computer Assisted Tomography) I and II trials, with blood samples taken at emergency department presentation (ROMICAT-I and -II) or at 2 and 4 h thereafter (ROMICAT-II). hs-cTn concentrations were measured using 3 assays (Roche Diagnostics, Elecsys 2010 platform; Abbott Diagnostics, ARCHITECT i2000SR; Siemens Diagnostics, HsVista). Per blood sample, we determined concordance across analytic benchmarks (
Results: Among 1,027 samples from 624 patients (52.8 ± 10.0 years; 39.4% women), samples were classified as
Conclusions: Caregivers should be aware that there are significant differences between hs-cTn assays in stratifying individual samples and patients with intermediate likelihood of ACS according to analytical benchmarks that may result in different management recommendations. (Rule Out Myocardial Infarction by Computer Assisted Tomography [ROMICAT]; NCT00990262) (Multicenter Study to Rule Out Myocardial Infarction by Cardiac Computed Tomography [ROMICAT-II]; NCT01084239).
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC8040768 | PMC |
http://dx.doi.org/10.1016/j.jacc.2021.01.046 | DOI Listing |