There is a plethora of severe acute respiratory syndrome-coronavirus-2 (SARS-CoV-2) serological tests based either on nucleocapsid phosphoprotein (N), S1-subunit of spike glycoprotein (S1) or receptor binding domain (RBD). Although these single-antigen based tests demonstrate high clinical performance, there is growing evidence regarding their limitations in epidemiological serosurveys. To address this, we developed a Luminex-based multiplex immunoassay that detects total antibodies (IgG/IgM/IgA) against the N, S1 and RBD antigens and used it to compare antibody responses in 1225 blood donors across Greece. Seroprevalence based on single-antigen readouts was strongly influenced by both the antigen type and cut-off value and ranged widely [0.8% (95% CI 0.4-1.5%)-7.5% (95% CI 6.0-8.9%)]. A multi-antigen approach requiring partial agreement between RBD and N or S1 readouts (RBD&N|S1 rule) was less affected by cut-off selection, resulting in robust seroprevalence estimation [0.6% (95% CI 0.3-1.1%)-1.2% (95% CI 0.7-2.0%)] and accurate identification of seroconverted individuals.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC7988055PMC
http://dx.doi.org/10.1038/s41598-021-86035-2DOI Listing

Publication Analysis

Top Keywords

accurate sars-cov-2
4
sars-cov-2 seroprevalence
4
seroprevalence surveys
4
surveys require
4
require robust
4
robust multi-antigen
4
multi-antigen assays
4
assays plethora
4
plethora severe
4
severe acute
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!