A Constant-Force End-Effector With Online Force Adjustment for Robotic Ultrasonography.

IEEE Robot Autom Lett

School of Biomedical Engineering and Imaging Sciences, King's College LondonKing's Health Partners, St Thomas' HospitalSE1 7EHLondonU.K.

Published: April 2021

In this letter, we propose a novel constant-force end-effector (CFEE) to address current limitations in robotic ultrasonography. The CFEE uses a parallel, motor-spring-based solution to precisely generate constant operating forces over a wide range and enable the ultrasound (US) probe to adapt to the abdominal contours autonomously. A displacement measurement unit was developed to realize the acquisition of probe position and precise control of the operating force. Moreover, the operating force can be adjusted online to maintain safety and continuity of operation. Simulations and experiments were carried out to evaluate the performance. Results show that the proposed CFEE can provide constant forces of 4-12 N with displacements of 0-8 mm. The maximum relative error of force generation is 8.28%, and the accuracy and precision for displacement measurement are 0.29 mm and ±0.16 mm, respectively. Various operating forces can be adjusted online during the same operation. Ultrasound images acquired by the proposed CFEE are of equally good quality compared to a manual sonographer scan. The proposed CFEE would have potential further medical applications.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC7968128PMC
http://dx.doi.org/10.1109/LRA.2021.3061329DOI Listing

Publication Analysis

Top Keywords

proposed cfee
12
constant-force end-effector
8
robotic ultrasonography
8
operating forces
8
displacement measurement
8
operating force
8
adjusted online
8
cfee
5
end-effector online
4