Biomechanical Comparison of Fibertape Device Repair Techniques of Ligamentous Lisfranc Injury in a Cadaveric Model.

Injury

The Pennsylvania State University, Center for Orthopaedic Research and Translational Science, 500 University Drive, H089 Hershey, PA, 17033. Electronic address:

Published: April 2021

Background: Lisfranc ligamentous injuries are complex, and their treatment, along with the preferred method of fixation, is controversial. Implementing a flexible synthetic augmentation device (fibertape) has been described as an alternative to traditional screw fixation. This biomechanical study evaluated two fibertape devices with interference screw fixation: InternalBrace, and InternalBrace with supplementary intercuneiform stabilization.

Methods: The diastasis and relative angular displacement between bones were measured at three midfoot joints in the Lisfranc articulation. Measurements were obtained for the pre-injured, injured, and post-fixation stages under static loading. Specimens then underwent stepwise increases in cyclic loading performed at 1 Hz and 100 cycles, at 100 N ground reaction force intervals from 500 to 1200 N to simulate postoperative loading, and then up to 1800 N to simulate high loads. Failure of fixation was defined as diastasis greater than 2 millimeters at the second-metatarsal - medial-cuneiform joint.

Results: InternalBrace specimens demonstrated failures in 3 of 9 (33%) specimens at cyclic loads of 1000 N. Conversely, InternalBrace with Supplementary Limb specimens had 1 failure at 1200 N. The difference in diastasis at the second metatarsal-medial cuneiform joint was statistically significant between the two groups at higher loads of 1600N (p = 0.019) and 1800N (p = 0.029).

Conclusion: The use of InternalBrace for ligamentous Lisfranc injuries appears to provide a biomechanically viable alternative for withstanding early postoperative protected weight bearing. Furthermore, the use of a supplementary limb in addition to the InternalBrace fibertape fixation method appears to enhance its biomechanical efficacy.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.injury.2021.02.077DOI Listing

Publication Analysis

Top Keywords

ligamentous lisfranc
8
screw fixation
8
internalbrace supplementary
8
supplementary limb
8
internalbrace
6
fixation
5
biomechanical comparison
4
fibertape
4
comparison fibertape
4
fibertape device
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!