Background: The aim of this study is to verify the analytical performance of four homocysteine detection systems made in China and to explore the comparability of homocysteine detection systems by isotope dilution liquid chromatography-tandem mass spectrometry (ID-LC-MS/MS) reference method.
Methods: The intra-batch precision, inter-batch precision, accuracy, and linear range of four homocysteine detection systems were evaluated. The ID-LC-MS/MS reference method was used to evaluate the comparability and accuracy of fresh frozen serum samples in four different detection systems of homocysteine. The ID-LC-MS/MS reference method is used to assign samples as calibrators to calibrate each system. The variation and deviation of fresh serum samples between different systems before and after calibration were compared.
Results: The intra-batch imprecision of the four detection systems was less than 5%, and the coefficient of variation of inter-batch imprecision was less than 6.7%. The precision met the clinical requirements. Before calibration, the results measured by detection system 2 are consistent with the ID-LC-MS/MS reference method, which meets the requirements of accuracy verification. The regression equation of R² ≥ 0.975 in the regression equation of linear analysis of the four systems, the linearity of the four detection systems is good in the range of evaluation concentration, and all of them can meet the declared linear range. The absolute average bias of fresh serum measured by the four detection systems after calibration decreased from 3.76 μmol/L, 0.96 μmol/L, 1.30 μmol/L, -1.56 μmol/L to 0.31 μmol/L, 0.28 μmol/L, 0.4 μmol/L, 0.40 μmol/L, respectively. The relative average bias decreased from 22.6%, 7.50%, 11.0% and -8.50% to 1.98%, 1.78%, 2.59%, 2.34%, respectively. After calibration, the slope and intercept of the regression curve of the fresh serum measured by the four detection systems and the reference method are closer to 1 and 0 than before calibration.
Conclusions: The precision, reference interval, and linear evaluation of the four detection systems are good. The ID-LC-MS/MS reference method assigning fresh frozen serum samples as calibrators can improve the accuracy and comparability of the results of different detection systems.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.7754/Clin.Lab.2020.200501 | DOI Listing |
JMIR Hum Factors
January 2025
New College of Florida, Sarasota, FL, United States.
Background: Bangladesh and West Bengal, India, are 2 densely populated South Asian neighboring regions with many socioeconomic and cultural similarities. In dealing with breast cancer (BC)-related issues, statistics show that people from these regions are having similar problems and fates. According to the Global Cancer Statistics 2020 and 2012 reports, for BC (particularly female BC), the age-standardized incidence rate is approximately 22 to 25 per 100,000 people, and the age-standardized mortality rate is approximately 11 to 13 per 100,000 for these areas.
View Article and Find Full Text PDFPLoS Negl Trop Dis
January 2025
Laboratorio de Ingeniería Genética y Biología Celular y Molecular-Área de virus de insectos, Departamento de Ciencia y Tecnología, Universidad Nacional de Quilmes, Quilmes, Buenos Aires, Argentina.
Mosquitoes are the primary vectors of arthropod-borne pathogens. Aedes aegypti is one of the most widespread mosquito species worldwide, responsible for transmitting diseases such as Dengue, Zika, and Chikungunya, among other medically significant viruses. Characterizing the array of viruses circulating in mosquitoes, particularly in Aedes aegypti, is a crucial tool for detecting and developing novel strategies to prevent arbovirus outbreaks.
View Article and Find Full Text PDFPLoS One
January 2025
North China Institute of Aerospace Engineering, Langfang, China.
As the global economy expands, waterway transportation has become increasingly crucial to the logistics sector. This growth presents both significant challenges and opportunities for enhancing the accuracy of ship detection and tracking through the application of artificial intelligence. This article introduces a multi-object tracking system designed for unmanned aerial vehicles (UAVs), utilizing the YOLOv7 and Deep SORT algorithms for detection and tracking, respectively.
View Article and Find Full Text PDFPLoS One
January 2025
College of Animal Science and Technology, Henan University of Science and Technology, Luoyang, China.
Background: As an opportunistic bacterial pathogen, Klebsiella pneumoniae (KP) is prone to causing a spectrum of diseases in rabbits when their immune system is compromised, which poses a threat to rabbit breeding industry. Bacillus coagulans (BC), recognized as an effective probiotic, confers a variety of benefits including anti-inflammatory and antioxidant properties.
Aim: The purpose of this study was to investigate whether dietary BC can effectively alleviate hepatic injury caused by KP.
PLoS One
January 2025
Department of Ocean Integrated Science, Chonnam National University, Yeosu, Korea.
Ensuring the supply of safe and high-quality drinking water can be compromised by the presence of chironomid larvae in drinking water treatment plants (DWTPs), which may contaminate municipal water systems through freshwater resources. Chironomids are dominant species known for their resilience to a broad range of extreme aquatic environments. This study aimed to identify the morphological characteristics and obtain genetic information of the chironomid Paratanytarsus grimmii found in the water intake source and freshwater resource of DWTPs in Korea, highlighting the potential possibility of a parthenogenetic chironomid outbreak within DWTP networks.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!