N-Acetylcysteine Interference with a Glucose Dehydrogenase Linked Glucose Meter.

J Diabetes Sci Technol

Laboratory Medicine, Division of Clinical Biochemistry, Saskatchewan Health Authority, Saskatoon, Saskatchewan, Canada.

Published: September 2022

Background: Our objective was to determine the effect of therapeutic concentrations of N-acetylcysteine, following intravenous infusion, on the measurement of blood glucose using a Roche Diagnostics glucose dehydrogenase-linked glucose meter compared to hospital laboratory methods.

Methods: N-acetylcysteine was added to aliquots of blood, with glucose promptly measured by the glucose meter, blood gas analyzer (glucose oxidase comparative method) and following centrifugation, plasma glucose measured with a hexokinase spectrophotometric comparative method. Glucose results were evaluated with linear regression and Bland Altman plots.

Results: In the presence of NAC, at concentrations greater than 5 mg/dL (0.31 mmol/L), positively biased glucose meter results were compared to the clinical laboratory results. Multivariate linear regression revealed that NAC-mediated meter results are influenced by NAC and glucose concentrations.

Conclusions: The addition of therapeutic concentrations of NAC to blood produces statistically significant positive biases when measured with the glucose dehydrogenase linked glucose meter device.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9445331PMC
http://dx.doi.org/10.1177/1932296821999416DOI Listing

Publication Analysis

Top Keywords

glucose meter
20
glucose
14
glucose dehydrogenase
8
dehydrogenase linked
8
linked glucose
8
therapeutic concentrations
8
blood glucose
8
meter compared
8
glucose measured
8
measured glucose
8

Similar Publications

To evaluate the accuracy of home self-monitoring portable blood glucose meters, we analyzed the current problems of patients using portable blood glucose meters and put forward reasonable suggestions. A self-designed questionnaire was used to survey 142 patients and 132 healthcare professionals. The questionnaire consisted of 16 items with an overall score ranging from 1 to 13 (with a higher score indicating better experience).

View Article and Find Full Text PDF

Aims: Physiological changes during pregnancy can infuence the performance of blood glucose meters. This study aimed to evaluate the analytical and clinical accuracy of glucose meters in pregnant women with hyperglycaemia.

Materials And Methods: Glucose was measured by four commonly used meters among consecutive women with diabetes in pregnancy.

View Article and Find Full Text PDF

Diagnostic Limitations of Applying a Human Portable Blood Glucose Meter in the Detection of Hypoglycemia in Pregnant Ewes.

Vet Sci

January 2025

Programa de Pós-Graduação em Ciências Veterinárias, Universidade Federal do Rio Grande do Sul (UFRGS), Porto Alegre 91540-000, Brazil.

Pregnant ewes are susceptible to hypoglycemia and ketosis; therefore, monitoring glycemic status is extremely important. Portable blood glucose meters (PBGMs) can assist in quickly and conveniently identifying glycemic disturbances in this species, provided that they meet the criteria of analytical accuracy. This study evaluated the performance of a human PBGM (Accu-Chek Performa, Roche Diagnostics, Basel, Switzerland) in the glycemic evaluation of 34 pregnant ewes at days 90 and 120 of pregnancy in comparison with the results of glycemia determination by a reference method (RM).

View Article and Find Full Text PDF

This study aims to determine and compare the reference values of the haematological and biochemical blood parameters of two fish species collected from the Gökova Bay (Muğla, South-Western of Türkiye): the non-native and invasive Randall's threadfin bream, and the native Common pandora, . Both species inhabit the same environment and compete for resources. Blood samples were collected from a total of 100 fish samples (50 and 50 ) which were caught from a depth of 30 to 60 meters between February 2023 and July 2024.

View Article and Find Full Text PDF

In Table 7.2, "Common interfering substances and/or conditions that affect glucose meters (for inpatient and outpatient use)," of the article cited above, the effects on glucose values measured by blood glucose meters for high and low hematocrit were incorrect. For high hematocrit, the effect would be falsely lower blood glucose values.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!