Aims: Previous studies have already shown early proximal ingrowth, fast osseous integration, and a stable fit of the uncemented Symax hip stem, with excellent clinical and radiographic performance. Aims were to evaluate cumulative revision rates and reasons for revision of the Symax hip stem using Dutch Arthroplasty Register (LROI) data and to assess possible associations between patient characteristics and revision rate of the Symax hip stem.
Patients And Methods: All total hip arthroplasties with the uncemented Symax hip stem registered in the LROI between 2007 and 2017 were included (n = 5,013). Kaplan-Meier survival analysis was performed to assess the cumulative 1, 5 and 7-year revision percentages. Cox proportional hazard regression analysis was performed to assess the association between patient and procedural characteristics, and revision arthroplasty of the stem.
Results: Cumulative 1, 5, and 7-year revision rates (with 95% confidence interval (CI)) for revision of any component were 1.5% (CI 1.2%-1.8%), 3.2% (CI 2.7%-3.7%), and 3.8% (CI 3.1%-4.4%) respectively. Cumulative 1, 5, and 7-year stem revision rates of the Symax hip stem were 0.9% (CI 0.6%-1.1%), 1.5% (CI 1.1%-1.9%), and 1.7% (CI 1.3%-2.1%) respectively. Periprosthetic fractures (n = 35) and loosening of the stem (n = 30) were the most common reasons for revision of the stem. Revision of the stem was associated with acute fracture as primary diagnosis (Hazard Ratio (HR) 2.4 (CI 1.3-4.3)), or history of a previous surgery to the affected hip (HR 2.7 (CI 1.4-5.2)).
Conclusion: This population-based registry study shows revision rates for the Symax hip stem comparable to those for best performing uncemented total hip arthroplasties in the Netherlands. Primary diagnosis of an acute fracture, and history of previous surgery on the affected hip, were significantly associated risk factors for revision of the Symax hip stem, and we discourage the use of the Symax hip stem in these patients.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC7954353 | PMC |
http://journals.plos.org/plosone/article?id=10.1371/journal.pone.0248483 | PLOS |
PLoS One
October 2021
Department of Orthopaedic Surgery, Research School Caphri, Maastricht University Medical Centre, Maastricht, The Netherlands.
Aims: Previous studies have already shown early proximal ingrowth, fast osseous integration, and a stable fit of the uncemented Symax hip stem, with excellent clinical and radiographic performance. Aims were to evaluate cumulative revision rates and reasons for revision of the Symax hip stem using Dutch Arthroplasty Register (LROI) data and to assess possible associations between patient characteristics and revision rate of the Symax hip stem.
Patients And Methods: All total hip arthroplasties with the uncemented Symax hip stem registered in the LROI between 2007 and 2017 were included (n = 5,013).
Acta Orthop
August 2020
Department of Orthopaedics, Leiden University Medical Center, Leiden, The Netherlands.
Acta Orthop
April 2020
Department of Orthopaedic Surgery, Research School Caphri, Maastricht University Medical Centre, Maastricht.
Background and purpose - The uncemented Symax hip stem has shown early proximal ingrowth as result of the BONIT-hydroxyapatite (HA) coating and the distal DOTIZE surface treatment. We evaluated 2-year postoperative radiostereometric analysis (RSA) migration of the Symax hip stem in THA patients. We also investigated the correlation between migration at 4 weeks and clinical outcomes after 2 years.
View Article and Find Full Text PDFJ Orthop Surg Res
July 2018
Department of Orthopaedic Surgery, Research School Caphri, Maastricht University Medical Centre, P. Debyelaan 25, P.O. Box 5800, 6202 AZ, Maastricht, the Netherlands.
Background: The uncemented Symax hip stem is developed through optimization of the uncemented Omnifit hip stem. The Symax stem design combines an anatomical anteverted proximal geometry with a straight distal section. The proximal part is coated with a biomimetic hydroxyapatite (HA) coating for improved osseointegration to enhance load transfer and to minimize proximal bone loss.
View Article and Find Full Text PDFHip Int
July 2018
Section for Hip and Knee Replacement, Department of Orthopaedic Surgery, Vejle Hospital, Vejle, Denmark.
Introduction: The design of the cementless Symax-HA femoral stem is based on geometrical analysis of human femoral anatomy to optimise the fit within the femur. The stem combines an anatomical proximal section enabling a metaphyseal anchorage with a straight distal section. This results in an improved proximal load transfer.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!