Cardiogenic shock is associated with high mortality. Patients often require temporary mechanical circulatory support. We aimed to develop a percutaneously implantable, assist device that unloads the left ventricle (LV) in a pulsatile way. The PERkutane KATheter pump technologie (PERKAT LV) device consists of a nitinol pump chamber, which is covered by foils carrying outflow valves. A flexible tube with a pigtail-shaped tip and inflow holes represents the distal part of the pump. The system is designed for 16F percutaneous implantation. The nitinol chamber is placed in the descending aorta while the flexible tube bypasses aortic arch and ascending aorta with its tip in the LV. An intra-aortic balloon pump is placed into the chamber and connected to a console. Balloon deflation generates a blood flow from the LV into the pump chamber. During balloon inflation, blood leaves the system through the outflow foil valves in the descending aorta. Under different afterload settings using a 30 cc intra-aortic balloon pump and varying inflation/deflations rates, we recorded flow rates up to 3.0 L/min. Based on this, we believe that PERKAT LV is a promising approach for temporary LV support. The proposed design and its excellent performance give basis for in vivo tests in an animal model.

Download full-text PDF

Source
http://dx.doi.org/10.1097/MAT.0000000000001397DOI Listing

Publication Analysis

Top Keywords

pump chamber
12
left ventricle
8
flexible tube
8
descending aorta
8
intra-aortic balloon
8
balloon pump
8
pump
6
concept design
4
design novel
4
novel pulsatile
4

Similar Publications

Introduction: Neurotrophic factors are widely known for their protective effect on spiral ganglion neurons (SGN) and the protection of these neurons is of great importance to optimize Cochlear Implants, which directly stimulate SGN in deaf patients. Previous studies have identified Cometin - also known as Meteroin-like - to be neuroprotective and beneficial for metabolic disorders. The aim of our study was to investigate the effects of different concentrations of recombinant human Cometin (hCometin) on SGN in regard to neuroprotection and neurite outgrowth and to evaluate its neurite guidance potential using a neurite outgrowth chamber.

View Article and Find Full Text PDF

Normothermic Crystalloid Polarizing Cardioplegia Improves Systolic and Diastolic Function in a Porcine Model of Cardiopulmonary Bypass.

Biomedicines

December 2024

Ludwig Boltzmann Institute for Cardiovascular Research, Center for Biomedical Research and Translational Surgery, Medical University of Vienna, 1090 Vienna, Austria.

Previously, we showed that blood-based polarizing cardioplegia exerted beneficial cardioprotection during hypothermic ischemia; however, these positive effects of blood-based polarizing cardioplegia were reduced during normothermic ischemia compared to blood-based hyperkalemic (depolarizing) cardioplegia. This study compares crystalloid polarizing cardioplegia to crystalloid depolarizing cardioplegia in a normothermic porcine model of cardiopulmonary bypass; Methods: Twelve pigs were randomized to receive either normothermic polarizing ( = 7) or depolarizing ( = 5) crystalloid cardioplegia. After the initiation of cardiopulmonary bypass, normothermic arrest (34 °C, 60 min) was followed by 60 min of on-pump and 90 min of off-pump reperfusion.

View Article and Find Full Text PDF

In coronary artery bypass grafting (CABG) on pump, achieving optimal visualization is critical for surgical precision and safety. The use of blowers to clear the CABG anastomosis poses risks, including the formation of micro-embolic gas bubbles, which can be insidious and increase the risk of cerebral or myocardial complications. This retrospective study compares the effectiveness of the use of irrigation mist and CO versus a direct CO blower without irrigation in terms of visualization, postoperative fibrillation, and micro-embolic gas activity.

View Article and Find Full Text PDF

The Heisenberg-RIXS instrument at the European XFEL.

J Synchrotron Radiat

January 2025

Institute Methods and Instrumentation for Synchrotron Radiation Research, Helmholtz-Zentrum Berlin für Materialien und Energie GmbH, Albert-Einstein-Straße 15, 12489 Berlin, Germany.

Resonant inelastic X-ray scattering (RIXS) is an ideal X-ray spectroscopy method to push the combination of energy and time resolutions to the Fourier transform ultimate limit, because it is unaffected by the core-hole lifetime energy broadening. Also, in pump-probe experiments the interaction time is made very short by the same core-hole lifetime. RIXS is very photon hungry so it takes great advantage from high-repetition-rate pulsed X-ray sources like the European XFEL.

View Article and Find Full Text PDF

Recapitulation of physiologic and pathophysiologic pulsatile CSF flow in purpose-built high-throughput hydrocephalus bioreactors.

Fluids Barriers CNS

December 2024

Department of Chemical Engineering and Materials Science, Wayne State University, 6135 Woodward Avenue, Rm 1413, Detroit, MI, 48202, USA.

Background: Hydrocephalus, an accumulation of cerebrospinal fluid (CSF) in the ventricles of the brain, is often treated via a shunt system to divert the excess CSF to a different compartment; if left untreated, it can lead to serious complications and permanent brain damage. It is estimated that one in every 500 people are born with hydrocephalus. Despite more than 60 years of concerted efforts, shunts still have the highest failure rate of any neurological device requiring follow-up shunt revision surgeries and contributing to the $2 billion cost of hydrocephalus care in the US alone.

View Article and Find Full Text PDF