PATIENT-SPECIFIC DOSE ESTIMATES IN DYNAMIC COMPUTED TOMOGRAPHY MYOCARDIAL PERFUSION EXAMINATION.

Radiat Prot Dosimetry

Department 6.2 Dosimetry for radiation therapy and diagnostic radiology, Physikalisch-Technische Bundesanstalt, Bundesallee 100, Braunschweig 38116, Germany.

Published: January 2021

The study aimed to implement realistic source models of a computed tomography (CT) scanner and Monte Carlo simulations to actual patient data and to calculate patient-specific organ and effective dose estimates for patients undergoing dynamic CT myocardial perfusion examinations. Source models including bowtie filter, tube output and x-ray spectra were determined for a dual-source Siemens Somatom Definition Flash scanner. Twenty CT angiography patient datasets were merged with a scaled International Commission on Radiological Protection (ICRP) 110 voxel phantom. Dose simulations were conducted with ImpactMC software. Effective dose estimates varied from 5.0 to 14.6 mSv for the 80 kV spectrum and from 8.9 to 24.7 mSv for the 100 kV spectrum. Significant differences in organ doses and effective doses between patients emphasise the need to use actual patient data merged with matched anthropomorphic anatomy in the dose simulations to achieve a reasonable level of accuracy in the dose estimation procedure.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC8227483PMC
http://dx.doi.org/10.1093/rpd/ncab016DOI Listing

Publication Analysis

Top Keywords

dose estimates
12
computed tomography
8
myocardial perfusion
8
source models
8
actual patient
8
patient data
8
effective dose
8
dose simulations
8
dose
5
patient-specific dose
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!