Introduction: We aim to describe the performance of combined IgM and IgG point-of-care antibody test (POC-Ab) (Wondfo®) compared to real-time reverse transcriptase (rRT-PCR) (Allplex™ 2019-nCoV Assay) in detecting coronavirus disease 2019 (COVID-19).

Methodology: We compared POC-Ab with rRT-PCR results among patients in a tertiary hospital from January to March 2020 in Bandung, Indonesia. We selected presumptive COVID-19 patients with positive rRT-PCR consecutively and 20 patients with negative rRT-PCR results were selected randomly from the same group of patients as controls. We described the sensitivity, specificity, positive predictive value (PPV), negative predictive value (NPV) with corresponding 95% confidence interval using serum and capillary blood samples. We also tested POC-Ab using non-COVID-19 (confirmed dengue and typhoid) patients' sera.

Results: Twenty-seven patients with positive rRT-PCR result and 20 negative controls were included (68.1% males, mean age 46 (SD: 15.4)). Using the serum, the sensitivity of the POC-Ab was 63.0% (42.4-80.6), specificity was 95.0% (75.1-99.9), PPV was 94.4% (72.7-99.8), NPV was 65.5% (45.7-82.1). A subset of 20 patients was tested using a capillary blood sample. The accuracy of the capillary blood sample is lower compared to serum (50.0% vs. 78.7%). None of the non-COVID-19 sera tested were reactive.

Conclusions: POC-Ab for COVID-19 has a high specificity with no false-positive result in non-COVID-19 sera. Therefore, it can be used to guide diagnostic among symptomatic patients in resource limited settings. Given its low sensitivity, patients with high suspicion of COVID-19 but non-reactive result should be prioritized for rRT-PCR testing.

Download full-text PDF

Source
http://dx.doi.org/10.3855/jidc.13370DOI Listing

Publication Analysis

Top Keywords

capillary blood
12
point-of-care antibody
8
antibody test
8
tertiary hospital
8
bandung indonesia
8
patients
8
patients positive
8
positive rrt-pcr
8
blood sample
8
non-covid-19 sera
8

Similar Publications

Among contributors to diffusible signaling are portal systems which join two capillary beds through connecting veins. Portal systems allow diffusible signals to be transported in high concentrations directly from one capillary bed to the other without dilution in the systemic circulation. Two portal systems have been identified in the brain.

View Article and Find Full Text PDF

Non-invasive, low intensity focused ultrasound is an emerging neuromodulation technique that offers the potential for precision, personalized therapy. An increasing body of research has identified mechanosensitive ion channels that can be modulated by FUS and support acute electrical activity in neurons. However, neuromodulatory effects that persist from hours to days have also been reported.

View Article and Find Full Text PDF

Over 46% of African pregnant women are anemic. Oral iron is recommended but often suboptimal, particularly late in pregnancy. Intravenous ferric carboxymaltose (FCM) could treat anemia in women in the third trimester in sub-Saharan Africa.

View Article and Find Full Text PDF

Purpose: To evaluate the presence and progression of maculopathy in patients with sickle cell disease (SCD) using Optical Coherence Tomography (OCT) and OCT-Angiography (OCTA), and to identify clinical/laboratory risk factors for progression during follow-up.

Methods: Complete ophthalmic examination, including fundoscopy and macular SD-OCT/OCTA scans, was performed in consecutive SCD-patients (HbSS/HbSβ0/HbSβ+/HbSC genotype) during baseline and follow-up visits. SCR stage was based on fundoscopy instead of the Goldberg classification, since fluorescein angiography was not routinely used.

View Article and Find Full Text PDF

Ischemia-reperfusion injuries are known to cause a range of retinal pathologies, including diabetic retinopathy, glaucoma, retinal vascular occlusions, and other vaso-occlusive conditions. This manuscript presents a method for inducing ischemia-reperfusion injury in a mouse model. The method utilized anterior chamber cannulation attached to a saline reservoir, generating hydrostatic pressure to raise the intraocular pressure to 90-100 mmHg.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!