Tumescent local anesthesia versus general anesthesia for subcutaneous implantable cardioverter-defibrillator implantation.

Heart Rhythm

Cardiac Arrhythmia Center, Montefiore-Einstein Center for Heart and Vascular Care, Division of Cardiology, Department of Medicine, Albert Einstein College of Medicine, New York, New York; Texas Cardiac Arrhythmia Institute, St. David's Medical Center, Austin, Texas. Electronic address:

Published: August 2021

Background: Subcutaneous implantable cardioverter-defibrillator (S-ICD) is an effective alternative to transvenous implantable cardioverter-defibrillator. General anesthesia (GA) is considered the standard sedation approach because of the pain caused by the manipulation of subcutaneous tissue with S-ICD implantation. However, GA carries several limitations, including additional risk of adverse events, prolonged in-room times, and increased costs.

Objective: The purpose of this study was to define the effectiveness and safety of tumescent local anesthesia (TLA) in comparison to GA in patients undergoing S-ICD implantation.

Methods: We performed a prospective, nonrandomized, controlled, multicenter study of patients referred for S-ICD implantation between 2019 and 2020. Patients were allocated to either TLA or GA on the basis of patient's preferences and/or anesthesia service availability. TLA was prepared using lidocaine, epinephrine, sodium bicarbonate, and sodium chloride. All patients provided written informed consent, and the institutional review board at each site provided approval for the study.

Results: Sixty patients underwent successful S-ICD implantation from July 2019 to November 2020. Thirty patients (50%) received TLA, and the rest GA. There were no differences between groups with regard to baseline characteristics. In-room and procedural times were significantly shorter with TLA (107.6 minutes vs 186 minutes; P < .0001 and 53.2 minutes vs 153.7 minutes; P < .0001, respectively). Pain was reported less frequently by patients who received TLA. The use of opioids was significantly reduced in patients who received TLA (23% vs 62%; P = .002).

Conclusion: TLA is an effective and safe alternative to GA in S-ICD implantation. The use of TLA is associated with shorter in-room and procedural times, less postprocedural pain, and reduced usage of opioids and acetaminophen for analgesia.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.hrthm.2021.03.008DOI Listing

Publication Analysis

Top Keywords

s-icd implantation
16
implantable cardioverter-defibrillator
12
received tla
12
tla
9
tumescent local
8
local anesthesia
8
general anesthesia
8
subcutaneous implantable
8
patients
8
in-room procedural
8

Similar Publications

Background: The subcutaneous implantable cardioverter-defibrillator (S-ICD) is an alternative to traditional ICDs. The PRAETORIAN score, based on chest radiographs, has been validated to predict the probability of successful S-ICD defibrillation testing by assessing factors like fat thickness between the coil and sternum and generator placement.

Objective: This study evaluated the correlation between the PRAETORIAN score and clinical characteristics, as well as implantation variables.

View Article and Find Full Text PDF

Antithrombotic therapy and the risk of pocket hematoma after subcutaneous implantable cardioverter-defibrillator implantation.

J Interv Card Electrophysiol

January 2025

Heart Center, Department of Cardiology, Amsterdam Cardiovascular Sciences Heart Failure & Arrhythmias, Amsterdam UMC Location University of Amsterdam, Meibergdreef 9, 1105, AZ, Amsterdam, the Netherlands.

Background: Little data exists regarding the optimal antithrombotic strategy during S-ICD implantation to prevent pocket hematomas. This study explores the association between perioperative antithrombotic management and the occurrence of pocket hematoma following S-ICD implantation.

Methods: All patients who underwent de novo S-ICD implantation between February 2009 and January 2023 at Amsterdam UMC were included.

View Article and Find Full Text PDF

Background: Noninferiority of omitting intraoperative defibrillation threshold (DFT) testing has been documented for transvenous implantable cardioverter defibrillators (ICD) whereas data for the subcutaneous-ICD (S-ICD) regarding the need for DFT testing, especially during S-ICD generator replacement, is not available.

Methods: A total of 112 consecutive patients who underwent S-ICD generator replacement and routine testing were included in this retrospective single-center study and analyzed regarding the outcome of intraoperative DFT.

Results: The majority of patients (87.

View Article and Find Full Text PDF

Unlabelled: New cardiac implantable electronic devices (CIEDs), such as leadless pacemakers and subcutaneous implantable cardioverter defibrillators (S-ICDs), are being used in patients with adult congenital heart disease. The selection of CIEDs often requires careful consideration due to technical challenges posed by a unique heart structure. A 27-year-old man following a surgical tetralogy of Fallot (TOF) repair developed non-sustained ventricular tachycardia, sick sinus syndrome, and complete atrioventricular block.

View Article and Find Full Text PDF

Inappropriate Therapy and Shock Rates Between the Subcutaneous and Transvenous Implantable Cardiac Defibrillator: A Secondary Analysis of the PRAETORIAN Trial.

Circ Arrhythm Electrophysiol

December 2024

Department of Cardiology, Amsterdam UMC Location University of Amsterdam, Heart Center, Amsterdam Cardiovascular Sciences Heart failure and Arrhythmias, the Netherlands (L.R.A.O.N., S.P., L.V.A.B., T.F.B., A.-F.B.E.Q., W.v.d.S., L.S., J.A.d.V., J.G.P.T., N.R.B., J.R.d.G., K.M.K., A.d.W., A.A.M.W., R.E.K.).

Article Synopsis
  • Inappropriate therapy (IAT) is a significant issue associated with implantable cardiac defibrillator (ICD) therapy, particularly highlighted in early subcutaneous ICD (S-ICD) studies which showed high rates of inappropriate shocks (IAS).
  • The PRAETORIAN trial, an international study with 849 patients, found no major differences in IAT and IAS rates between S-ICD and transvenous ICD (TV-ICD) groups, as both groups had similar cumulative incidences.
  • Key predictors for IAT varied between the two groups, with TV-ICD patients experiencing IAT mainly from supraventricular tachycardias, while S-ICD patients faced issues from cardiac oversensing
View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!